Wilcoxon REF2500 Operation Manual Download Page 4

 

97029  Rev C.1  09/18 

 

Page 4 of 15 

 

1.0 Introduction

 

 

This guide is designed to assist the user in the proper operation of the ReferenceMate portable 
reference source (handheld shaker).  Further information is provided on the use of accessory items. 

 

2.0 Description 
 

The ReferenceMate portable vibration reference source enables users in the field to easily 
verify sensor performance and the integrity of the cabling between the sensor and read-out 
equipment, either online or portable units.  
  
A built-in reference accelerometer assures that a 1 g test level is maintained for a unit under 
test (sensor and mounting hardware) weighing up to 250 grams. 
 
Frequency of operation and measurement types (peak/RMS) are selected with front panel push 
buttons. 
 
•  61.4 Hz for imperial measurements, where 1 g is equal to 1 inch per second (ips), ideal for 

use with velocity sensors 

•  100.0 Hz to confirm accelerometer performance at the same reference frequency used by 

most manufacturers 

•  159.2 Hz for metric measurements, where 1 g is equal to 9.81 m/sec

2

 and 9.81 mm/sec 

 

A bottom mounted D-ring belt clip is included for hands-free transport between 
locations when in the field. The belt clip can be removed and replaced with a 70 lbf 
pull force magnet (optional) for attachment to metal structures for hands-free 
operation. 
 
LEDs notify the user when the battery condition falls below operational levels or if 
the unit is overloaded. Reference test points are provided to test and check 
calibration of the shaker’s internal accelerometer. 

 

3.0 Applicable models 

 

The manual covers the model REF2500 handheld shaker included in the REF2510 and REF2510R 
kits, and the use of applicable accessories, both standard and optional. 

 
 
 
 
 
 
 
 
 
 
 
 
 

 

Summary of Contents for REF2500

Page 1: ...rive Frederick MD 21701 USA Amphenol Maryland Inc d b a Wilcoxon Sensing Technologies 97029 Rev C 1 09 18 Tel 1 301 330 8811 Tel 1 800 WILCOXON Fax 1 301 330 8873 www wilcoxon com ReferenceMate portab...

Page 2: ...uld be consulted and read carefully CAUTION This symbol is used when caution is needed to prevent damage to equipment It is used where careful attention to certain procedures described in the instruct...

Page 3: ...ower ON OFF 9 5 2 3 Frequency selection 10 5 2 4 RMS PEAK 10 5 2 5 BAT low power LED 11 5 2 6 OOR LED 11 5 2 7Orientation during operation 11 6 0 Testing triaxial transducers 12 6 1 Universal AC adapt...

Page 4: ...z for imperial measurements where 1 g is equal to 1 inch per second ips ideal for use with velocity sensors 100 0 Hz to confirm accelerometer performance at the same reference frequency used by most m...

Page 5: ...Remove the battery holder and insert four AA batteries into the holder being careful to observe correct polarity see figure 1B After the batteries are installed insert the holder into the battery cup...

Page 6: ...and any mounting accessory is mounted to the ReferenceMate head connect all cables from the UUT to the readout equipment You are now ready to test your device 3 Press the ON OFF button to initiate the...

Page 7: ...n the ReferenceMate off before the end of its autoshutdown cycle Press the ON OFF button Figure 2 control panel of the ReferenceMate 5 2 Operational details Peak RMS mode indicators Indicate peak or R...

Page 8: ...n RMS The equivalent peak to peak displacement values in mils are shown in Table 1a for reference For those using the metric system Table 1b shows the equivalent values All the values below are repres...

Page 9: ...712 sensors 24 20 2 7 2 3 13249 02 1 4 28 to 10 32 adaptor plate 24 18 2 7 2 0 13267 01 90 adaptor plate 993 series sensors 24 24 2 7 2 7 SF6M 1 1 4 28 to M6 adaptor stud 24 30 2 7 3 4 SF6M 1 4 28 to...

Page 10: ...l units US One g at 61 4 Hz is equal to 1 in sec ips velocity 100 Hz is the preferred frequency for checking accelerometer performance against calibration data 100 Hz is the most common reference freq...

Page 11: ...ed OOR LED will be illuminated during the control time cycle Once the LED goes out the unit is ready for operation This is normally less than 3 seconds 5 2 7 Orientation during operation The Reference...

Page 12: ...eration of the ReferenceMate2500 is required part number REF003 is available This AC adaptor comes with several universal plug adaptors and will operate from 50 60 Hz 110 220 VAC To power the shaker w...

Page 13: ...the test point voltage should be done only in an ESD safe environment Figure 5 Test points 7 1 Background information The REF2500 contains an internal sine wave generator with a reference acceleromet...

Page 14: ...f high levels of distortion an incorrect output level or out of tolerance frequency is observed the REF2500 should be returned to the factory for service There are no user serviceable parts in the uni...

Page 15: ...ack cover must be securely installed on the unit Do not expose the unit to dust or moisture without both of these protective mechanisms in place 8 0 Storage If the unit is going to be out of service f...

Reviews: