9
WIKA operating instructions models SC58, SC64
14054366.02 02/2016 EN/DE
EN
3. Specifications
Specifications
Model SC58
Model SC64
Nominal size in mm
Front dimensions: 62 x 28
Mounting dimensions:
58.2 x 25.2
60
Sensor element
Pt100 resistance thermometer (standard), Pt1000 or PTC
(selectable via software menu)
Measuring range
Pt100, Pt1000: -80 … +400 °C
PTC:
-50 … +130 °C
Measuring accuracy
±1 K or ±0.5 % of measuring range (greater value applies)
The actual value display can be an integer or have a
resolution of 0.5 K or 0.1 K
Display
3-digit LED display, 13 mm high, colour red
Power supply
■
Model SCxx.230
■
Model SCxx.024
AC 230 V, 50/60 Hz, current
consumption max. 20 mA
AC 12 ... 24 V, DC 12 ... 36 V,
current consumption max.
30 mA
AC 230 V, 50/60 Hz, current
consumption max. 20 mA
AC 12 ... 24 V, DC 16 ... 36 V,
current consumption max.
30 mA
Switching output
Relays 12 A resistive load,
(2.2 A inductive load), 250 V,
change over contact
Relays 16 A resistive load,
(2.2 A inductive load), 250 V,
contact normally open
Electrical
connections
Screw terminals for cable up
to 1.5 mm
2
Pin 1 up to 4: Fast-on
connection, 6.3 x 0.8 mm
Pin 5 and 6: Fast-on
connection, 2.8 x 0.5 mm
Case
Plastic, black
Weight
approx. 200 g
Ingress protection
Front panel IP50 (panel
mounting)
Terminals IP00
Front panel IP65 (panel
mounting)
Terminals IP00
Mounting option
■
Model SC58
■
Model SC64
Snap-in mounting
Panel mounting with mounting bracket
3. Specifications