WIKA operating instructions models LSO.06, LSO.25
31
13334603.01 09/2010 GB/D/F/E
GB
Replacing fuse of switching amplifier
■
Switch net supply off
■
Taking fuse out
Use a new fuse only with the value printed on the circuit board
Supply voltage
Model LSO.06
AC 230 V
T 50 mA
AC 115/120 V
T 100 mA
AC 24 V
T 400 mA
DC 24 V with potential separation
see marking on circuit board
DC 24 V without potential separation
T 400 mA
■
Switch net supply on
At spare part orders indicate absolutely exact type, series no. and
commission no. the manufacturer.
9.2 Cleaning
If the glass tip becomes soiled, clean it carefully.
CAUTION!
■
Before cleaning, correctly disconnect the instrument from the pressure
supply, switch it off and disconnect it from the mains.
■
Clean the instrument with a moist cloth.
■
Electrical connections must not come into contact with moisture.
■
Wash or clean the dismounted instrument before returning it, in order to
protect staff and the environment from exposure to residual media.
■
Residual media in dismounted instruments can result in a risk to persons,
the environment and equipment. Take sufficient precautionary measures.
For information on returning the instrument see chapter "12.2 Return".
9. Maintenance, repair and cleaning