12
WIKA operating instructions, models F4801, F4802, F4818
ADPR1X914110.01 08/2020 EN/DE
EN
=
=
=
=
5. Commissioning, operation
5.3 Mounting the single point load cell
Single point load cell
Mounting surface
Platform
Spacer part
Spring displacement
■
Screw the underside of the single point load cell to the mounting surface and
make sure that the opposite side of the single point load cell does not rest on the
mounting surface (no force shunt).
■
If necessary, use a spacer part . The thickness of the spacer part corresponds to
the spring displacement .
■
Align the platform so that the centre of the platform corresponds with the centre of
the single point load cell .
■
Screw the upper side of the single point load cell to the platform and make sure
that the opposite side of the single point load cell has no contact with the platform
(no force shunt).
■
If necessary, use a spacer part .