EN
60
WIKA operating instructions, model CTR3000
14191761.01 05/2019 EN/DE
Button
Result
Download selected
Downloads all screenshots marked with a
[
]
to the USB memory stick
Delete selected
Deletes all screenshots marked with a
[
]
from the instrument
The downloaded screenshot is a file in the format *.png.
Name of a screenshot in the directory *CTR3000\Screenshots:
Ddmmyyyy_hhmmss
The set time and data format can change according on the programming in the application
[Settings]
.
6.5.3 Calibration files
By pressing the button
[Calibration files]
a dialogue opens on the right side. All created calibration files are listed there then
on the left side.
To select or delete these please use the right button.
Button
Result
Select all
Ticks all calibration files with a
[
]
in the box in front of the calibration files name
Deselect all
Removes the
[
]
in the box in front of the calibration files name
Download selected
Downloads all log files marked with a
[
]
to the USB memory stick
The downloaded calibration file is a file in the format *.xml and also *.csv.
Following information is stored in the file in the directory *CTR3000\CalibrationFiles:
■
Routine name
■
Start and end time
■
Stability criteria
■
Used equipment
■
Calibration results: set value, reference value, reference stability, DUT value, DUT raw value
Name of a screenshot in the directory *CTR3000\CalibrationFiles:
Routine_Ddmmyyyy_hhmm
The set time and data format can change according on the programming in the application
[Settings]
.
More information about the content of the downloaded calibration file see chapter 6.4.6.3 “Start the routine”.