WIKA operating instructions, models CTH6300 and CTH6500
EN
31
2079988.08 02/2018 EN/DE/IT
6.
Use the ▲▼ arrow keys to select the desired calibration function.
Standard characteristic curve oFF
1.
Use the ▲▼ arrow keys to select
oFF
.
2. Press
ENTER/MENU
to confirm the desired setting.
3. Use
ESC
to return to the measuring mode.
Calibration by code oP1
1.
Use the ▲▼ arrow keys to select
oP1
.
2. Press
ENTER/MENU
to confirm the desired setting.
⇒
A small
1
appears in the lower part of the display. 4 characters follow it (hex code
/ 0 ... F).
3.
Change the 4 characters using the ▲ arrow key.
4.
Press the ▼ arrow key to go to the next position.
5. When all 4 characters are entered as required, confirm by pressing
ENTER/MENU
.
⇒
A small
2
appears and the following 4 characters can be changed as well.
6. Use
ESC
to return to the measuring mode.
6.4.5 Activating/deactivating measuring channels [Chnl] (only for 2-channel
instruments)
Chnl = Channel = select
1. Press the
ENTER/MENU
key and select
Chnl
using the ▲▼ arrow keys.
2. Press the
ENTER/MENU
key once again.
⇒
A small
1
appears on the left side of the display, it indicates the channel.
3.
Use the ▲▼ arrow keys to select the channel to be activated or deactivated.
4. Confirm the selection by pressing
ENTER/MENU
.
CAL
CoFF
oP1
oP2
6. Commissioning, operation