WIKA CTD9100-1100 Operating Instructions Manual Download Page 11

11

WIKA operating instructions, model CTD9100-1100

14030725.02 08/2019 EN/DE

EN

5. Transport, packaging and storage / 6. Commissioning, operation

5.  Transport, packaging and storage

5.1  Transport

Check the temperature dry-well calibrator for any damage 

that may have been caused by transport.

Obvious damage must be reported immediately.

CAUTION!

Damage through improper transport

With improper transport, a high level of damage 

to property can occur.

 

When unloading packed goods upon delivery 

as well as during internal transport, proceed 

carefully and observe the symbols on the 

packaging.

 

With internal transport, observe the 

instructions in chapter 5.2 “Packaging and 

storage”.

If the instrument is transported from a cold into a warm 

environment, the formation of condensation may result 

in instrument malfunction. Before putting it back into 

operation, wait for the instrument temperature and the room 

temperature to equalise.

5.2  Packaging and storage

Do not remove packaging until just before mounting.

Keep the packaging as it will provide optimum protection 

during transport (e.g. change in installation site, sending for 

repair).

Permissible conditions at the place of storage:

 

Storage temperature: -10 ... +60 °C [14 ... 140 °F]

 

Humidity: 30 ... 95 % relative humidity (no condensation)

Avoid exposure to the following factors:

 

Direct sunlight or proximity to hot objects

 

Mechanical vibration, mechanical shock (putting it down 

hard)

 

Soot, vapour, dust and corrosive gases

 

Potentially explosive environments, flammable 

atmospheres

Store the temperature dry-well calibrator in its original 

packaging in a location that fulfils the conditions listed above.

6.  Commissioning, operation

Personnel:

 Skilled personnel

Protective equipment:

 Protective gloves

Only use original parts (see chapter 13 “Accessories”).

WARNING!

Physical injuries and damage to property 

and the environment caused by hazardous 

media

Upon contact with hazardous media (e.g. 

oxygen, acetylene, flammable or toxic 

substances), harmful media (e.g. corrosive, 

toxic, carcinogenic, radioactive), and also with 

refrigeration plants and compressors, there is 

a danger of physical injuries and damage to 

property and the environment.

Should a failure occur, aggressive media and/

or with high temperature may be present at the 

instrument.

 

For these media, in addition to all standard 

regulations, the appropriate existing codes or 

regulations must also be followed.

 

Wear the requisite protective equipment.

6.1  Operating position

The operating position of the temperature dry-well calibrator 

is in the vertical orientation, since this guarantees an optimal 

temperature distribution in the metal block.

The calibrator is only for indoor use.

Voltage supply

DANGER!

Danger to life caused by electric current

Upon contact with live parts, there is a direct 

danger to life.

 

Only ever use the supplied mains cable (see 

chapter 4.5 “Voltage supply”).

The voltage supply of the instrument is made via the power 

cable. This is included in the scope of delivery.

6.2  Switching on the calibrator

1.  Connect to the mains using the mains connector supplied.

2.  Switch on the mains switch.

Summary of Contents for CTD9100-1100

Page 1: ...Operating instructions Betriebsanleitung EN DE Temperature dry well calibrator model CTD9100 1100 Temperatur Blockkalibrator Typ CTD9100 1100 Temperature dry well calibrator model CTD9100 1100...

Page 2: ...nguages can be found at www wika com 04 2011 WIKA Alexander Wiegand SE Co KG All rights reserved Alle Rechte vorbehalten WIKA is a registered trademark in various countries WIKA ist eine gesch tzte Ma...

Page 3: ...oning operation 11 6 1 Operating position 11 6 2 Switching on the calibrator 11 6 3 Start up procedure 12 6 4 Temperature switch test 12 6 5 Testing of temperature sensors 12 6 6 Positioning of the pr...

Page 4: ...icinity of the instrument and readily accessible to skilled personnel at any time Pass the operating instructions onto the next operator or owner of the instrument Skilled personnel must have carefull...

Page 5: ...r 12 Specifications The appropriate instrument should be selected depending on the application it should be connected correctly tests carried out and all components serviced This instrument is not per...

Page 6: ...instrument If the calibrator is used for purposes not expressly specified in these operating instructions additional safety measures have to be taken The Processor regulator has been configured in fac...

Page 7: ...ific regulations current standards and directives are capable of carrying out the work described and independently recognising potential hazards Special operating conditions require further appropriat...

Page 8: ...art contains the complete electronic unit for controlling the reference temperature Solid state relays SSR are used to control the heating and cooling elements Constant use at extreme temperatures red...

Page 9: ...creasing the value shown on display 2 Holding down the button will decrease the speed 7 Button Increasing the value shown on display 2 Holding down the button will increase the speed 8 Function menu b...

Page 10: ...cable are the scope of delivery The cable with the CEE polarized plug is designed to reduce the risk of electric shock polarizing the phase of the power If possible always use these cable Rx Tx 13 14...

Page 11: ...ases Potentially explosive environments flammable atmospheres Store the temperature dry well calibrator in its original packaging in a location that fulfils the conditions listed above 6 Commissioning...

Page 12: ...there is evidence of the status of the test item This ensured that the test item requires a short time until it reaches the temperature of the metal block 6 6 Positioning of the probe The sleeve is ma...

Page 13: ...button M This opens the main menu 2 With the button M to select the desired entry in the main menu see overview 3 With the button E confirm the selected menu item The calibrator has three function lev...

Page 14: ...14 WIKA operating instructions model CTD9100 1100 14030725 02 08 2019 EN DE EN 7 2 1 Menu structure parameter levels Please note Button F button M Menu structure 7 Operating of the calibrator...

Page 15: ...rs and on the lower the set point is displayed Set point 2 is reached acoustic signal and the ramp is paused A cooling ramp is set gradient must be negative and the set point 2 must be lower than the...

Page 16: ...then the buttons M and have to be pressed simultaneously Back over simultaneous pressing of the buttons M and or waiting 20 s Function Meaning Access key Access code Default 2 Baud rate Velocity of d...

Page 17: ...vironment caused by high temperatures Should a failure with high temperature may be present at the instrument Wear the requisite protective equipment CAUTION Physical injuries and damage to property a...

Page 18: ...al media in the dismounted instrument can result in a risk to personnel the environment and equipment Use the requisite protective equipment Carry out the cleaning process as described below 1 Cool do...

Page 19: ...medium Clean the instrument in order to protect persons and the environment from exposure to residual media 11 1 Dismounting WARNING Risk of burns During dismounting there is a risk of high temperatu...

Page 20: ...m 1 67 x 2 56 in Insert material Ceramic Voltage supply Power supply AC 230 V 50 60 Hz or AC 110 V 50 60 Hz Power consumption 950 VA Fuse 6 3 A slow blow fuse Power cord AC 230 V for Europe Communicat...

Page 21: ...21 WIKA operating instructions model CTD9100 1100 14030725 02 08 2019 EN DE EN 170 6 69 448 17 64 330 12 99 390 15 35 12 Specifications Top view Side view left Rear view Front view Dimensions in mm in...

Page 22: ...13 Accessories Accessories Order code Description CTX A K9 Transport case TB Power cord for EU EU for Switzerland CH for UK UK for USA Canada US Insert replacement tool RT Ordering information for you...

Page 23: ...Lagerung 31 6 Inbetriebnahme Betrieb 31 6 1 Betriebslage 31 6 2 Einschalten des Kalibrators 31 6 3 Anfahrprozedur 32 6 4 Temperaturschaltertest 32 6 5 Pr fen von Temperaturf hlern 32 6 6 Positionierun...

Page 24: ...he des Ger tes f r das Fachpersonal jederzeit zug nglich aufbewahrt werden Betriebsanlei tung an nachfolgende Benutzer oder Besitzer des Ger tes weitergeben 1 Allgemeines Das Fachpersonal muss die Be...

Page 25: ...rfen keinesfalls berschritten werden siehe Kapitel 12 Technische Daten Es ist je nach Anwendungsfall ein entsprechendes Instru ment auszuw hlen dieses korrekt anzuschlie en Tests durchzuf hren sowie...

Page 26: ...egelger t konzipiert Bei einem Gebrauch des Kalibrators f r nicht ausdr cklich in dieser Betriebsanleitung vorgesehene Anwendungen m ssen zus tzliche Schutzma nahmen getroffen werden Der elektronische...

Page 27: ...sowie Kenntnis der landesspezifischen Vorschriften geltenden Normen und Richtlinien in der Lage die beschriebenen Arbeiten auszuf hren und m gliche Gefahren selbstst ndig zu erkennen Spezielle Einsatz...

Page 28: ...lette Elektronik einheit zur Regelung der Referenztemperatur Zur Ansteuerung der Heizelemente werden Halbleiter Relais SSR verwendet Konstante Verwendung des Kalibrators an extre men Temperaturen verr...

Page 29: ...Wert welcher auf Anzeige 2 dargestellt wird Bei Halten der Taste wird die Geschwindigkeit verringert 7 Taste Erh hen des Wertes welcher auf Anzeige 2 darge stellt wird Bei Halten der Taste wird die G...

Page 30: ...in Euro Steckerkabel beinhaltet Das CEE Kabel wird mitgeliefert um einen elektrischen Schlag aufgrund Verpolung zu verhindern Wenn m glich immer diese Kabel benutzen Rx Tx 13 14 15 16 17 Bei Benutzung...

Page 31: ...ellen Ru Dampf Staub und korrosive Gase Explosionsgef hrdete Umgebung entz ndliche Atmosph ren Temperatur Blockkalibrator in der Originalverpackung an einem Ort lagern der die oben gelisteten Bedingun...

Page 32: ...nn eine Aussage ber den Zustand des Pr flings gemacht werden Hierbei darauf achten dass der Pr fling kurze Zeit ben tigt bis er die Temperatur des Metallblockes angenommen hat 6 6 Positionierung der T...

Page 33: ...7 Bedienung des Kalibrators 7 1 Einstellen einer tempor ren Soll Temperatur Sollwertmodus Einstellung der Soll Temperatur Dr cken der Taste erh ht den Sollwert Dr cken der Taste erniedrigt den Sollwe...

Page 34: ...34 WIKA Betriebsanleitung Typ CTD9100 1100 14030725 02 08 2019 EN DE DE 7 Bedienung des Kalibrators 7 2 1 Men struktur Parameterebenen Bitte beachten Taste F Taste M Men struktur...

Page 35: ...ert auf der zweiten Zeile des Displays angezeigt Ist der Sollwert 2 erreicht ert nt ein Signal und die Rampe wird automatisch pausiert Wenn eine runter laufende Rampe eingestellt wurde muss der Gradie...

Page 36: ...Code ber eingestellt wurde und danach gleichzeitig die Tasten M und gedr ckt werden Zur ck kommt man durch gleichzeitiges Dr cken der Tasten M und oder durch 20 s langes Warten Funktion Bedeutung Acce...

Page 37: ...hutzhandschuhe WARNUNG K rperverletzungen Sach und Umweltsch den durch hohe Temperaturen Am Ger t k nnen im Fehlerfall extreme Tempe raturen anliegen Notwendige Schutzausr stung tragen VORSICHT K rper...

Page 38: ...SICHT Besch digung des Ger tes Eine unsachgem e Reinigung f hrt zur Besch digung des Ger tes Keine aggressiven Reinigungsmittel verwen den Keine harten und spitzen Gegenst nde zur Reinigung verwenden...

Page 39: ...Um Sch den zu vermeiden 1 Das Ger t mit dem D mmmaterial in der Verpackung platzieren Zu allen Seiten der Transportverpackung gleichm ig d mmen 2 Wenn m glich einen Beutel mit Trocknungsmittel der Ver...

Page 40: ...7 x 2 56 in H lsenmaterial Keramik Spannungsversorgung Hilfsenergie AC 230 V 50 60 Hz oder AC 110 V 50 60 Hz Leistungsaufnahme 950 VA Sicherung 6 3 A tr ge Netzkabel AC 230 V f r Europa Kommunikation...

Page 41: ...anleitung Typ CTD9100 1100 14030725 02 08 2019 EN DE DE 12 Technische Daten 170 6 69 448 17 64 330 12 99 390 15 35 Abmessungen in mm in Ansicht von oben Ansicht von der Seite links Ansicht von hinten...

Page 42: ...13 Zubeh r Zubeh r Bestellcode Beschreibung CTX A K9 Transportkoffer TB Netzkabel f r die EU EU f r die Schweiz CH f r UK UK f r USA Kanada US H lsenwechselwerkzeug RT Bestellangaben f r Ihre Anfrage...

Page 43: ...43 WIKA operating instructions model CTD9100 1100 14030725 02 08 2019 EN DE...

Page 44: ...Alexander Wiegand SE Co KG Alexander Wiegand Stra e 30 63911 Klingenberg Germany Tel 49 9372 132 0 Fax 49 9372 132 406 info wika de www wika de 44 WIKA subsidiaries worldwide can be found online at w...

Reviews: