background image

15

WIKA operating instructions process pressure transducer, model CPT6020

0020507001F  05/2019 EN

EN

5.1.4  Electrical Connections

Note:

 Power supply should be 

9 to 18 VDC (12 VDC nominal). 

The pinouts on the six pin connector  and RS-232 and RS-485 Cable connections are:

 

External Connector Pinouts

Pin

Function

Pigtail Color

1

RS-485 TX-

White

2

+9-18V

Brown

3

Ground

Gray

4

RS-485 RX+ or RS-232 RX

Black

5

RS-485 RX-

Blue

6

RS-485 TX+ or RS-232 TX

Pink

HOST

DB9

COM

Female

GND

PWR

12 VDC

GRAY

BLACK

PINK

2
3

5

BROWN

RS-232

HOST

DB9

COM

Female

GND

PWR

12 VDC

GRAY

BLACK

PINK

2
3

5

BROWN

RS-485

9

1

BLUE

WHITE

Figure 5.1.2 COM Connection Pin-out and RS-232 / RS-485 cable connections

Installation

Summary of Contents for CPT6020

Page 1: ...Operating instructions Precision Pressure Transducer Basic Version Model CPT6020 EN Precision Pressure Transducer Model CPT6020 ...

Page 2: ...Operating instructions model CPT6020 12 2017 Mensor LP All rights reserved Mensor is a registered trademark of Mensor LP All other brand and product names are trademarks or registered trademarks of their respective companies Prior to starting any work read the operating instructions Keep for later use ...

Page 3: ... 3 7 Labeling Safety Marks 11 3 7 1 Product Label 11 3 7 2 Symbols 12 3 8 Warnings and Cautions 12 4 Transport Packaging and Storage 13 4 1 Transport 13 4 2 Packaging and storage 13 5 Installation 14 5 1 Mounting 14 5 1 1 Pressure Connections 14 5 1 2 Safety Relief Valves 14 5 1 3 Pressure Media 14 5 1 4 Electrical Connections 15 6 Operation 16 6 1 General Operation 16 6 1 1 Command and Query Form...

Page 4: ...4 7 2 2 Environment 25 7 2 3 Pressure Standards 25 7 2 4 Media 25 7 2 5 Setup 25 7 2 6 Calibration Communication Software 26 8 Dismounting Return and Disposal 28 8 1 Dismounting 28 8 2 Return 29 8 3 Disposal 29 9 Specifications 30 9 1 Measurement Specification 30 9 2 General Specification 31 9 3 Approvals and Certificates 31 9 4 Dimensions in mm in 32 10 Accessories 33 10 1 Fittings Gaskets 33 10 ...

Page 5: ...5 WIKA operating instructions process pressure transducer model CPT6020 0020507001F 05 2019 EN EN ...

Page 6: ... modifications Factory calibrations are carried out in accordance with international standards Further information Mensor Corporation Address 201 Barnes Dr San Marcos TX 68666 Internet address www mensor com Relevant data sheet CT 25 12 Application consultant Tel 1 512 396 4200 1 800 984 4200 USA only Fax 1 512 396 1820 E Mail sales mensor com techservices mensor com Importer for Europe WIKA Alexa...

Page 7: ...eration in industrial residential or commercial environments 1 3 Software License Agreement This product contains intellectual property i e software programs that are licensed for use by the end user customer hereinafter end user This is not a sale of such intellectual property The end user shall not copy disassemble or reverse compile the software program The software programs are provided to the...

Page 8: ...s as low as 10 H2O to as high as 15 015 psia Figure 2 1 2 1 Features Here is a short list of significant features designed into the CPT6020 Accuracy to 0 02 FS Wide voltage power input range IP 67 rating for harsh environments All 316L SS wetted path on ranges 5 psi Command emulation for digital communications User selectable units Compact size 2 2 Scope of Delivery Precision Pressure Transducer m...

Page 9: ...on 3 2 Intended Use The CPT6020 pressure transducer is designed to measure pressure with extreme accuracy as stand alone devices or as incorporated in OEM type applications It is characterized over a wide temperature range and is designed to be used in environments that are much more harsh than tolerable by previous generations of Mensor pressure transducers 3 3 Improper Use WARNING Injuries throu...

Page 10: ...le injury and damage to equipment The activities described in these operating instructions may only be carried out by skilled personnel who have the qualifications described below Skilled personnel Skilled personnel authorized by the operator are understood to be personnel who based on their technical training knowledge of measurement and control technology and on their experience and knowledge of...

Page 11: ...tructions displayed in the work area regarding personal protective equipment The requisite personal protective equipment must be provided by the operating company Wear safety goggles Protect eyes from flying particles and liquid splashes Wear a protective helmet Protects the head from falling objects Wear safety shoes Protect feet from falling objects or objects lying around as well as against tox...

Page 12: ...een trained in proper safety practices WARNING NOT EXPLOSION PROOF Installation of this instrument in an area requiring devices rated as intrinsically safe is not recommended WARNING POSSIBLE INJURY The tubing valves and other apparatus attached to the gauge must be adequate for the maximum pressure which will be applied otherwise physical injury to the operator or bystanders is possible CAUTION U...

Page 13: ... the symbols on the packaging If the instrument is transported from a cold into a warm environment the formation of condensation may result in instrument malfunction Before putting it back into operation wait for the instrument temperature and the room temperature to equalize 4 2 Packaging and storage Permissible conditions at the place of storage Storage Temperature 40 85 C Humidity 5 95 relative...

Page 14: ... Damage to the Instrument When working on open electrical circuits printed circuit boards there is a risk of damaging sensitive electronic components through electrostatic discharge The correct use of grounded working surfaces and personal armbands is required 5 1 Mounting The transducer may be mounted via its process connection or with any ring clamps along its body Having the sensor oriented in ...

Page 15: ...5 Cable connections are External Connector Pinouts Pin Function Pigtail Color 1 RS 485 TX White 2 9 18V Brown 3 Ground Gray 4 RS 485 RX or RS 232 RX Black 5 RS 485 RX Blue 6 RS 485 TX or RS 232 TX Pink HOST DB9 COM Female GND PWR 12 VDC GRAY BLACK PINK 2 3 5 BROWN RS 232 HOST DB9 COM Female GND PWR 12 VDC GRAY BLACK PINK 2 3 5 BROWN RS 485 9 1 BLUE WHITE Figure 5 1 2 COM Connection Pin out and RS ...

Page 16: ...ormat On RS 232 versions of the CPT6020 the or address character is not necessary to precede the command for the Sensor command set But the Mensor Legacy command set still requires a proceeding and address character With the default sensor command set in RS 232 the start character and address character are not required 6 1 1 3 Command Error Detection and Debugging The CPT6020 has built in response...

Page 17: ...r lf All commands that are password protected are designated with Password Protected in the Query column below Global Commands Query Command Data Description N A ADDRESS sp C RS 485 only C Transducer address 0 9 A Z IDN ID N A Manu Model SN SW Version Manufacturer Model SN SW Version N A SAVE Saves Data to EEPROM TYPE N A C Pressure type 1 char FILTER FILTER sp n n Filter percentage 1 to 99 WINDOW...

Page 18: ...sp n n Configures output for PRESS Query and Continuous Output see 6 1 2 10 Table 6 1 2 Sensor Command Set Password Protection In the Sensor Command Set data that affects calibration is protected by a user settable 4 character password designated as Password Protected The default password is 0000 This is changed using the PWD_CHANGE command It should be changed once the user receives the device Th...

Page 19: ...the following parameters to the shown settings Parameter Setting Filter 90 Window 10 Baud 57600 CMD_SET 0 CUST_UNIT 1 Table 6 1 2 2 Default Settings 6 1 2 2 Setting the Command Set CMD_SET This will set the transducer to communicate using different command sets Command Set Description 0 Sensor Default Command Set 1 Mensor Legacy Table 6 1 2 3 Command Set List ...

Page 20: ...t Used 12 ftSW 0C 32 uHg 0C 13 atm 33 tsi 14 bar 34 mHg 0C 15 mbar 35 hPa 16 mmH2O 4C 36 Mpa 17 cmH2O 4C 37 mmH2O 20C 18 MH2O 4C 38 cmH2O 20C 19 mmHg 0C 39 mH2O 20C 20 cmHg 0C 99 CUST_UNIT Table 6 1 2 5 Available Unit Codes and Units 6 1 2 4 Customer Units The CUST_UNIT multiplier in conjunction with UNIT_INDEX 99 allows the user to set their own units The CUST_UNIT is a multiplier times psi to ac...

Page 21: ...e selected parameters will populate the output string with commas separating the variables in the order in which they appear in the following table Parameter Description Weight Format Max Length plus 1 comma between fields Pressure Pressure in current units Fixed n nnnnnnnE nn 14 Units Unit Description 1 CCCCCCCCCC 10 Stable 0 1 unstable stable 16 n 1 Error 0 1 no errors error s in stack 32 n 1 Ch...

Page 22: ...T sp Y R Change CPT6020 into different command sets 1 is Mensor Legacy 0 is Sensor command set None DC XDC X sp DC sp Y Returns the date of calibration where Y is mmddyy Cust DC Y XDC sp Y R Loads the date of calibration where Y is mmddyy None DT XDT X sp DT sp NNN Returns the temperature a d DAC offset None FL XFL X sp FL sp NN Returns the filter it is the percentage of old reading added to new r...

Page 23: ...rrection multiplier None T XT X sp T sp C Returns pressure type None W XW X sp W sp NN Returns the filter window in FS Mens W NN XW sp NN R Sets the filter window in counts Readings inside this window are filtered None ZC XZC X sp ZC sp Y Returns customer zero correction value in current units Cust ZC X XZC sp Y R Sets customer zero correction value in current units Table 6 1 3 Mensor Legacy Comma...

Page 24: ...es launch to the announcement of discontinuation typically 8 to 10 years All repairs should be performed by Mensor due to the complexity of performing these repairs 7 2 Recalibration The CPT6020 automatically calculates the pressure reading for all effects of temperature and non linearity within the compensated temperature range The Process is referred to as dynamic compensation because each readi...

Page 25: ... nitrogen or clean dry instrument air For ranges above 5 psi any liquid media compatible with 316L SS is acceptable A reference height variation between the standard and the CPT6020 can cause errors A calculation should be made to compensate for this difference 7 2 5 Setup The schematic in Figure 7 2 5 shows a typical setup used to calibrate the CPT6020 Deadweight Tester or Primary Standard Power ...

Page 26: ...tion of the transducer output and configuration of various parameters can be accomplished It has many features that allow you to read and adjust the rich feature set including Changing filter settings Modifying the output format of the sensor Changing the pressure units output It also has many features to adjust various calibration parameters after the appropriate password is sent Setting the cali...

Page 27: ...27 WIKA operating instructions process pressure transducer model CPT6020 0020507001F 05 2019 EN EN Maintenance and Recalibration ...

Page 28: ...n plants and compressors there is a danger of physical injuries and damage to property and the environment Before storage of the dismounted instrument following use wash or clean it in order to protect persons and the environment from exposure to residual media Wear the requisite protective equipment see Chapter 3 6 Personal protective equipment Observe the information in the material safety data ...

Page 29: ...truments with lithium ion rechargeable batteries or lithium metal batteries The included lithium ion rechargeable batteries or lithium metal batteries are subject to the requirements of the dangerous Goods Act For shipping special packaging and labelling requirements must be observed A dangerous goods expert must be consulted when preparing the item to be shipped Do not send any damaged or defecti...

Page 30: ...or reserves the right to change specifications without notice 9 1 Measurement Specification Accuracy 1 0 020 FS Measuring ranges Gauge pressure 0 0 36 to 0 1500 psig 0 25 mbar to 0 100 bar Bi directional pressure 2 0 18 to 15 1500 psig 12 5 mbar to 1 100 bar Absolute pressure 0 5 psia to 0 15 015 psia 0 350 mbar to 0 1001 bar abs Calibration interval 185 days Pressure units 39 and 1 user defined C...

Page 31: ...rrosive gases Ranges 5 psi media compatible with the listed wetted parts Voltage supply Power supply 9 to 18 VDC 12 VDC nominal Power consumption 26 mA at 12 VDC 5 Permissible ambient conditions Compensated temperature range 0 to 50 oC 32 to 122 oF Operating temperature range 40 to 85 oC 40 to 185 oF Storage temperature range 40 to 85 oC 40 to 185 oF Humidity 0 95 r h non condensing Operating alti...

Page 32: ...108 9mm 4 29in 85 7mm 3 37in 18 9mm 74in 31 8mm 1 25in 6 PIN M8 CONNECTOR 1 16 BARBED REFERENCE PORT 1500 psi 1500 to 6000 psi 6000 to 15000 psi SAE J514 37 FLARE PORT 7 16 20 THREAD 117 6mm 4 63in 85 7mm 3 37in 18 8mm 74in 31 8mm 1 25in 6 PIN M8 CONNECTOR SEAL SCREW SAE J514 37 FLARE PORT 7 16 20 THREAD 122 9mm 4 84in 25 3mm 1 00in 31 8mm 1 25in 95 0mm 3 74in 6 PIN M8 CONNECTOR SEAL SCREW AUTOCLA...

Page 33: ...vailable fittings to connect to the transducer have the following Mensor Part numbers Part Number Description 6000604023 Male NPT 6000604024 7 16 20 SAE 6000604025 BSPP 4250040006 Copper gasket washer Many other fittings are available consult the factory if you need assistance 10 2 Accessories Part Number Description 4050060021 Communication Cable 0020528001 Power Supply Communication Cable Access...

Page 34: ...xander Wiegand SE Co KG Alexander Wiegand Straße 30 63911 Klingenberg Germany Tel 49 9372 132 0 Fax 49 9372 132 406 info wika de www wika de Mensor LP 201 Barnes Drive San Marcos TX 78666 USA Tel 1 512 3964200 15 Fax 1 512 3961820 E Mail sales mensor com www mensor com WIKA subsidiaries worldwide can be found online at www wika com ...

Reviews: