10. Specifications
10.3 Model CPH6200 hand-held pressure indicator (complete measuring chain)
Model CPH6200 hand-held pressure indicator (complete measuring chain)
Accuracy of the
measuring chain
1)
■
0.2 % FS
■
0.1 % FS at reference conditions
2)
(not for pressure ranges
< 100 mbar [< 1.45 psi])
Mean temperature
coefficient
≤ 0.2 % of span/10 K (outside the reference conditions)
2)
Compensated range
0 ... 80 °C [0 ... 176 °F]
1)
It is defined by the total measurement uncertainty, which is expressed with the coverage factor (k = 2) and includes the following
factors: the intrinsic performance of the instrument, the measurement uncertainty of the reference instrument, long-term stability,
influence of ambient conditions, drift and temperature effects over the compensated range during a periodic zero point correction.
2) Reference conditions: 15 ... 25 °C [59 ... 77 °F]
10.4 Approvals
CE conformity
Model CPH6200
■
EMC directive
EN 61326 emission (group 1, class B) and immunity (portable
equipment)
■
RoHS directive
Model CPT6200
■
EMC directive
EN 61326 emission (group 1, class B) and immunity (portable
measuring equipment)
■
Pressure equipment directive
Module A, internal production control
■
RoHS directive
For further approvals, see website
10.5 Certificates
Certificate
Calibration for
model CPT6200
■
3.1 inspection certificate per EN 10204 (factory calibration)
■
DAkkS calibration certificate (traceable and accredited in
accordance with ISO/IEC 17025)
Recommended
calibration interval
1 year (dependent on conditions of use)
For certificates, see website
For further specifications, see WIKA data sheet CT 11.01 and the order documentation.