WIKA operating instructions, model CPH6200
EN
11221780.03 07/2022 EN/DE
Pos. Symbol Meaning
⇒
The adjustment can only be carried out if the deviation is less than
500 digits. If a zero point correction has been made, this will be
signalled by the message '
nuLL
Corr
' being displayed briefly as the
instrument is switched on.
■
Restoring the factory calibration
By pressing the
[TARA]
button for approx. 15 seconds, the factory settings
will be restored.
10
QUIT
STORE
Activate hold function or logger function
⇒
■
Hold function
By pressing the
[STORE/QUIT]
button, the last measured value will be
shown in the lower display. Pressing it again hides the value again (only if
logger is deactivated).
■
Logger function
Activation via the
[STORE/QUIT]
button, only if the logger function has
been selected via the main menu, see chapter 6.5 “Data logger”.
11
▲
MIN
Display of the respective Min. memory value
By pressing the
[MIN]
button, the minimum value measured will be displayed.
Pressing it again hides it.
To clear the Min. memory, press the
[MIN]
button for > 2 seconds.
12
MENU
SET
Enter configuration
By pressing the
[SET/MENU]
button for approx. 2 seconds, the settings
such as configuration, adjustment, alarm, logger and system clock can be
accessed.
■
Differentiation
By pressing the
[SET/MENU]
button, the secondary display will show the
difference of channel 1 to channel 2 (DIF = CH1 - CH2). Pressing it again
undoes this action. (Only with 2-channel version and 2 connected pressure
sensors).
Abbreviations, definitions
“XXX”
Menu XXX is accessed
[XXX]
Press button XXX
‚XXX‘
Display of a message 'XXX'
4.3 Electrical connections
On the upper edge of the instrument are located the connection sockets CH1 and CH2
(CH2 only with the 2-channel version) for the connection of model CPT6200 pressure
sensors, as is the socket for the connection of the WIKA interface cable, see chapter
4.5 “Pressure sensors” and chapter 4.6 “Serial interface”.
4. Design and function