5. Commissioning, operation

Once the cover of the case has been removed, to configure the instrument 

continue as follows:

 

To access the menu, while the actual value is displayed, press key 2 for 2 

seconds until “

dP

” appears in the display.

 

To set the parameter values, press keys 2 and 3.

 

To save the set value, press key 1. The parameter name will then appear in 

the display once more. The parameter name will then appear in the display 

once more.

 

To switch to the next parameter, press key 1. The name of the parameter 

will appear in the display.

If, during input, no key is pressed for 60 seconds, the config

-

uration of the instrument will be interrupted. Previously saved 

values will not be lost.

Parameters of the A-IAI-1 attachable indicator

Parameters Values

Meaning

Key 1

Keys 2 and 3

dP

Position of the decimal point (decimal point)

----

Max. indication range: -1999 … 9999

---.-

Max. indication range: -199.9 … 999.9

--.--

Max. indication range: -19.99 … 99.99

-.---

Max. indication range: -1.999 … 9.999

di.Lo

Lower indication range limit (display low)

-1999 ... 9999 This value will be displayed when the input 

signal = 4 mA

di.Hi

Upper indication range limit (display high)

-1999 ... 9999 This value will be displayed when the input 

signal = 20 mA

Summary of Contents for A-IAI-1

Page 1: ...Operating instructions Betriebsanleitung DE EN Attachable indicator model A IAI 1 Aufsteckanzeige Typ A IAI 1 Attachable indicator model A IAI 1...

Page 2: ...r model A IAI 1 2 2012 WIKA Alexander Wiegand SE Co KG All rights reserved Alle Rechte vorbehalten WIKA is a registered trademark in various countries WIKA ist eine gesch tzte Marke in verschiedenen L...

Page 3: ...ns for instruments per ATEX 8 3 5 Special hazards 9 3 6 Labelling safety marks 11 3 7 Ex marking 11 4 Transport packaging and storage 12 4 1 Transport 12 4 2 Packaging and storage 12 5 Commissioning o...

Page 4: ...egulations for the instrument s range of use The operating instructions are part of the product and must be kept in the immediate vicinity of the instrument and readily accessible to skilled personnel...

Page 5: ...external aids through the direct input of the upper and lower measuring range limits and the decimal point position The parameters and limit values are entered using three keys All programmable param...

Page 6: ...ctions 3 1 Explanation of symbols WARNING indicates a potentially dangerous situation that can result in serious injury or death if not avoided CAUTION indicates a potentially dangerous situation that...

Page 7: ...ugs and sockets must be protected from contamination If the instrument is transported from a cold into a warm environment the formation of condensation may result in instrument malfunction Before putt...

Page 8: ...result in the loss of explosion protection Electrical specifications in accordance with the EC type examination certificate Input circuit In Ex ia ib IIC IIB intrinsic safety ignition protection Only...

Page 9: ...e exami nation certificate and the relevant country specific regulations for installation and use in hazardous areas e g IEC 60079 14 NEC CEC Non observance can result in serious injury and or damage...

Page 10: ...es e g a PC Under certain circumstanc es internal connections in third party devices e g GND connected to the safety earth can lead to impermissible voltages If the connected cable is longer than 30 m...

Page 11: ...ct label example Model Approval related data Supply voltage Order code Date of manufacture Serial number Symbols Before mounting and commissioning the instrument ensure you read the operating instruct...

Page 12: ...vibration mechanical shock putting it down hard Soot vapour dust and corrosive gases Hazardous environments flammable atmospheres WARNING Physical injuries and damage to property and the environment I...

Page 13: ...ect connection could damage the indicator CAUTION Note the maximum permissible input current of 40 mA 5 2 Pin assignment The assignment of the angular connector is designed for the most commonly used...

Page 14: ...ge the wires from contact 1 and contact 2 so that these match the connection of the transmitter to be attached Now wire both contacts in the angular connector accordingly General information on changi...

Page 15: ...this task Once the configuration has been completed correctly place the seal and fit the cover The display is optionally available with the membrane keys on the front of the instrument The configurati...

Page 16: ...tch to the next parameter press key 1 The name of the parameter will appear in the display If during input no key is pressed for 60 seconds the config uration of the instrument will be interrupted Pre...

Page 17: ...uring range limits The measuring range is limited exactly to the input signal When this exceeds or drops below it the indication range limit is displayed e g for a humidity of 0 100 r h by dropping be...

Page 18: ...the display once more To switch to the next parameter press key 1 The name of the parameter will appear in the display Parame ters Values Meaning Key 1 Keys 2 and 3 oFFS Zero offset offset 5 00 5 00 T...

Page 19: ...in value will be shown To recall Max value Press key 2 briefly Hi will be displayed briefly and then for approx 2 s the Max value will be shown To delete the Min Max values Press keys 2 and 3 simulta...

Page 20: ...ng incorrect The error message will reset itself once the display value is once more within the permis sible limits Err 7 System error Permissible operating temperature too high or too low Instrument...

Page 21: ...lead to damage to the instrument Do not use any aggressive cleaning agents Do not use any hard or pointed objects for cleaning 8 Dismounting return and disposal WARNING Physical injuries and damage to...

Page 22: ...m 2 Place the instrument along with shock absorbent material in the packag ing Place shock absorbent material evenly on all sides of the transport packaging 3 If possible place a bag containing a desi...

Page 23: ...ical connection To intrinsically safe transmitter with 4 20 mA output and angular connector per DIN 175301 803 A Supply voltage Not required since the attachable indicator is powered by the 4 20 mA lo...

Page 24: ...n the indication 0 1 10 K Safety related maximum values Supply voltage Ui DC 9 28 V Short circuit current Ii 100 mA Power limitation Pi 1 200 mW Maximum effective internal capacitance Ci 13 nF Maximum...

Page 25: ...EX 30 3 5 Besondere Gefahren 31 3 6 Beschilderung Sicherheitskennzeichnungen 33 3 7 Ex Kennzeichnung 33 4 Transport Verpackung und Lagerung 34 4 1 Transport 34 4 2 Verpackung und Lagerung 34 5 Inbetri...

Page 26: ...chen Unfallverh tungsvorschriften und allgemeinen Sicherheitsbestimmungen einhalten Die Betriebsanleitung ist Produktbestandteil und muss in unmittelbarer N he des Ger tes f r das Fachpersonal jederze...

Page 27: ...lgt ohne externe Hilfsmittel durch direkte Eingabe der oberen und unteren Messbereichsgrenze und der Dezimal punktposition Die Parameter und Grenzwerte werden ber drei Tasten eingegeben Alle programmi...

Page 28: ...ation hin die zum Tod oder zu schweren Verletzungen f hren kann wenn sie nicht gemieden wird VORSICHT weist auf eine m glicherweise gef hrliche Situation hin die zu geringf gigen oder leichten Verletz...

Page 29: ...Pr zisionsmessger te mit erforderlicher Sorgfalt behandeln vor N sse St en starken Magnetfeldern statischer Elektrizit t und extremen Temperaturen sch tzen keine Gegenst nde in das Ger t bzw ffnungen...

Page 30: ...schriften zur Unfallverh tung erf llen 3 4 Zus tzliche Sicherheitshinweise f r Ger te nach ATEX WARNUNG Die Nichtbeachtung dieser Inhalte und Anweisungen kann zum Verlust des Explosionsschutzes f hren...

Page 31: ...reinigt werden die eine elektrostatische Aufladung erzeugen k nnen 3 5 Besondere Gefahren WARNUNG Die Angaben der geltenden Baumusterpr fbescheinigung sowie die jeweiligen landesspezifischen Vorschrif...

Page 32: ...r t und Anschl ssen alle Teile gegen direktes Ber hren gesch tzt sind Die Beschaltung besonders sorgf ltig beim Anschluss an andere Ger te z B PC konzipieren Unter Umst nden k nnen interne Verbindunge...

Page 33: ...ngen Typenschild Beispiel Typ Zulassungsrelevante Daten Hilfsenergie Bestellcode Herstellungsdatum Seriennummer Symbole Vor Montage und Inbetriebnahme des Ger tes unbedingt die Betriebsanleitung lesen...

Page 34: ...he Vibration mechanischer Schock hartes Aufstellen Ru Dampf Staub und korrosive Gase Explosionsgef hrdete Umgebung entz ndliche Atmosph ren WARNUNG K rperverletzungen Sach und Umweltsch den Eine unsac...

Page 35: ...olgen Bei falschem Anschluss kann das Anzeigeger t zerst rt werden VORSICHT Unbedingt den max zul ssigen Eingangsstrom von 40 mA beachten 5 2 Anschlussbelegung Die Winkelsteckerbelegung ist auf die ge...

Page 36: ...epasst werden Hierzu den Winkelstecker der A IAI 1 ffnen und die Dr hte von Kontakt 1 und Kontakt 2 so austauschen dass diese dem Anschluss des anzuschlie enden Trans mitters entsprechen Nun noch die...

Page 37: ...igung von geeigneten ESD Schutzma nahmen durchzuf hren Offen liegende Bauteile bei dieser Arbeit nicht besch digen Nach Beendigung der Konfiguration die Dichtung korrekt einle gen und den Deckel monti...

Page 38: ...n Der Name des Parameters erscheint im Display Wird bei der Eingabe l nger als 60 Sekunden keine Taste gedr ckt so wird die Konfiguration des Ger tes abgebrochen Bereits gespeicherte Werte gehen nicht...

Page 39: ...Anzeige Messbereichsgrenze Messbereich ist genau auf das Eingangs signal begrenzt Bei ber Unterschreitung wird die Display Anzeigebereichsgrenze angezeigt z B f r Feuchte 0 100 r F bei Unter berschrei...

Page 40: ...n chsten Parameter zu wechseln Taste 1 dr cken Der Name des Parameters erscheint im Display Parameter Werte Bedeutung Taste 1 Tasten 2 und 3 oFFS Nullpunktverschiebung offset 5 00 5 00 Die Eingabe de...

Page 41: ...s der Min Wert angezeigt Abruf des Max Wertes Taste 2 kurz dr cken es wird kurz Hi und anschlie end f r ca 2 s der Max Wert angezeigt L schen des Min Max Wertes Taste 2 und 3 gleichzeitig f r 2 s dr c...

Page 42: ...ierung fehlerhaft Die Fehlermeldung wird zur ckgesetzt sobald der Anzeigewert wieder innerhalb der zugelassenen Grenzen liegt Err 7 Systemfehler zul ssige Betrieb stemperatur ber bzw unterschritten Ge...

Page 43: ...m e Reinigung f hrt zur Besch digung des Ger tes Keine aggressiven Reinigungsmittel verwenden Keine harten und spitzen Gegenst nde zur Reinigung verwenden 8 Demontage R cksendung und Entsorgung WARNUN...

Page 44: ...nh llen 2 Das Ger t mit dem D mmmaterial in der Verpackung platzieren Zu allen Seiten der Transportverpackung gleichm ig d mmen 3 Wenn m glich einen Beutel mit Trocknungsmittel der Verpackung beif gen...

Page 45: ...her Anschluss An eigensicheren Messumformer mit Ausgang 4 20 mA und Winkelstecker nach DIN 175301 803 A Hilfsenergie Nicht erforderlich da die Aufsteckanzeige aus der 4 20 mA Schleife versorgt wird Sp...

Page 46: ...icherheitstechnische H chstwerte Hilfsenergie Ui DC 9 28 V Kurzschlussstrom Ii 100 mA Leistungsbeschr nkung Pi 1 200 mW Maximale wirksame innere Kapazit t Ci 13 nF Maximale wirksame innere Induktivit...

Page 47: ...WIKA Betriebsanleitung Aufsteckanzeige Typ A IAI 1 47 DE 14016825 04 07 2021 EN DE Anlage 1 EU Konformit tserkl rung Annex 1 EU declaration of conformity...

Page 48: ...1 48 WIKA Alexander Wiegand SE Co KG Alexander Wiegand Strasse 30 63911 Klingenberg Germany Tel 49 9372 132 0 Fax 49 9372 132 406 info wika de www wika de WIKA subsidiaries worldwide can be found onl...

Reviews: