EN
5.3 Mounting
WARNING!
Damage to property and the environment through
unsuitable sealing materials
Hot media might escape if unsuitable sealing material is
used. This can cause damage to the plant and pollution
of the environment.
▶
Use a sealing material that is suitable for the
temperature range, and that is resistant to the medium
being measured.
Before mounting
Before starting the work, make sure that the conditions for
mounting the thermometer are met:
■
The plant is disconnected and free from pressure.
■
The plant has cooled down to ambient temperature.
■
There is a male thread at a suitable position in the plant.
■
A suitable sealing material is used.
Only use suitable, correctly sized tools for mounting.
5. Commissioning, operation