![Widex USB LINK User Instructions Download Page 24](http://html1.mh-extra.com/html/widex/usb-link/usb-link_user-instructions_979162024.webp)
24
Electromagnetic immunity
The USB Link is intended for use in the electromagnetic environment specified below. The cus-
tomer or the user of a USB Link should assure that it is used in such an environment.
Immunity Test
IEC 60601
Test level
Compliance
level
Electromagnetic
environment – guidance
Electrostatic
discharge (ESd)
IEc 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, con-
crete or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be
at least 30 %.
Electrical fast
transients/burst
IEc 61000-4-4
± 2 kV for
power line sup-
plies
± 1 kV for input/
output lines
not
applicable *)
not
applicable *)
not applicable *)
Surge
IEc 61000-4-5
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
not
applicable *)
not
applicable *)
not applicable *)
Voltage dips,
short interrup-
tions and volt-
age variations
on power supply
input lines
IEc 61000-4-11
<5 %
U
T
(>95 %
dip in
U
T
) for
0.5 cycle
40 %
U
T
(60 % dip in
U
T
)
for 5 cycles
70 %
U
T
(30 % dip in
U
T
)
for 25 cycles
<5 %
U
T
(>95 % dip
in
U
T
) for 5 s
not
applicable *)
not applicable *)
power frequen-
cy (50/60 Hz)
magnetic field
IEc 61000-4-8
3 A/m
3 A/m
power frequency magnetic
fields should be at levels charac-
teristic of a typical location
in a typical commercial or
hospital environment
noTE
U
T
is the a.c. mains voltage prior to the application of the test level.
*) Equipment powered via USB cable
Summary of Contents for USB LINK
Page 1: ...1 Widex USB Link User instructions ...
Page 6: ...6 4 When the installation is completed select Finish ...
Page 9: ...9 3 Select Finish and the Widex USB Link is ready to be used ...
Page 34: ...34 ...
Page 35: ...35 ...
Page 36: ...36 CIB199 0411 2011 04 9 502 1810 041 01 Nymoellevej 6 DK 3540 Lynge Denmark www widexpro com ...