Immunity Test
IEC 60601
Test level
Compliance
level
Electromagnetic
environment –
guidance
Electrostatic dis-
charge (ESD)
IEC
61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV con-
tact
± 8 kV air
Floors should be
wood, concrete or
ceramic tile. If floors
are covered with
synthetic material,
the relative humidi-
ty should be at least
30%.
Electrical fast tran-
sients/burst
IEC
61000-4-4
± 2 kV for power line
supplies
± 1 kV for input/out-
put lines
Not
applicable *)
Not
applicable *)
Not
applicable *)
Surge
IEC
61000-4-5
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
Not
applicable *)
Not
applicable *)
Not
applicable *)
31