Electrical fast
transients/burst
IEC 61000-4-4
± 2 kV for
power line
supplies
± 1 kV for input/
output lines
TV-BASE:
± 2 kV for power
line supplies
± 1 kV for input/
output lines
TV-CONTROLLER:
Not applicable *)
Mains power quality
should be that of a
typical commercial or
hospital environment .
Surge
IEC 61000-4-5
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
TV-BASE:
± 1 kV line(s) to
line(s)
±2kV (live to earth
& neutral to earth)
Combination
wave
TV-CONTROLLER:
Not applicable *)
Mains power quality
should be that of a
typical commercial or
hospital environment .
53
Summary of Contents for TV-DEX
Page 1: ...USERINSTRUCTIONS TV DEX ...
Page 2: ...2 ...
Page 5: ...PACKAGECONTENTS User manual TV Controller Box 1 Mini USB power supply 5 ...
Page 15: ...4a 3a POWER TV AUDIO 15 ...
Page 17: ...3b 4b 17 ...
Page 19: ...7 Place the controller in the base and allow it to charge for eight hours the first time 19 ...
Page 24: ...WARNING The breakaway buckle of the lanyard must be positioned at the back of your neck 1 24 ...
Page 31: ...WARNING Do not attempt to change the battery Contact your hearing care professional 31 ...
Page 73: ...73 ...
Page 74: ...74 ...
Page 75: ...75 ...