Whitehall SP-1600-A-220 Operation And Maintenance Manual Download Page 7

WHITEHALL MANUFACTURING

 • P.O. BOX 3527 • City of Industry, CA 91744-0527 U.S.A

Phone (800) 782-7706 • (626) 968-6681 • Fax (626) 855-4862 • Web: www.whitehallmfg.com

5

NOTE:  If  the  appliance  is  operated  without  water,  the 
concealed heating element may be subjected to damage 
and may no longer heat up the tank even though power 
is  "ON".  If  heating  element  becomes  damaged,  it  will 
have to be replaced.

--------------------------------------

I filled my Water Bath with water, plugged it in, turned 
it  "ON"  (power  switch  is  not  illuminated),  the 
Temperature Controller display is not illuminated, and 
the unit is not heating up the water.

Mains  circuit  breaker  may  have  "tripped".  Reset  the 
power mains circuit breaker. Plug the Water Bath into a 
G.F.C.I.  protected  electrical  receptacle.  Do  not  plug  in 
any other devices or appliances to the same receptacle 
the Water Bath is using.

Equipment  Circuit  Protection  fuse(s)  may  have  "blown". 
Check fuse(s) for electrical continuity using a Multi-Meter 
(combination  Volt,  Ampere,  &  Ohm  meter).  Replace 
fuse(s) if necessary. Be sure to use the correct fuse size 
& type. Refer to the Repair Parts section of this manual.

Electrical  component(s)  such  as  power  inlet,  fuse 
holders,  fuse  carriers,  power  switch,  quick  disconnect 
terminals,  wire  connectors,  wiring,  etc.  may  have 
become  faulty.  A  qualified  electrical  technician  should 
perform  an  electrical  continuity  inspection  of  the  entire 
electrical system using a Multi-Meter to identify possibly 
bad  electrical  component(s).  Replace  electrical 
component(s) as required.

--------------------------------------

My Water Bath temperature raises and lowers by itself?

The  Water  Bath  is  provided  with  an  Over-Temp 
thermostat.  If  the  operating  thermostat  fails  the 
temperature  will  rise.  When  a  temperature  of  102°C 
(215°F)  is  achieved  the  over-temp  thermostat  will  open 
the circuit and the temperature will drop to 82°C (180°F), 
then  rise  again.  This  cycle  will  continue  until  the 
operating thermostat is replaced.

The  appliance  will  require  a  qualified  technician  to  turn 
over  the  unit;  carefully  remove  the  bottom  cover  by 
removing  the  mounting  screws.  The  technician  will 
troubleshoot and/or replace the operating thermostat. 

The  heating  element  should  be  visually  and  electrically 
inspected to ensure that it is still in good, proper working 
condition.  Make  sure  the  heating  element  did  not 
delaminate  from  the  underside  of  the  tank.  Make  sure 
the  heating  element  still  has  electrical  continuity  by 
checking  circuit  with  a  Multi-Meter.  Check  that  the 
heating  element  does  not  have  dark  discolored  spots, 
which may indicate that the element has burned.

Before  reassembling,  be  sure  to  perform  a  visual  and 
electrical  inspection  of  the  rest  of  the  electrical  system. 
Make certain all other electrical components are in good, 
working condition. Replace any components that appear 
to be burned or have become faulty.

--------------------------------------

How do I adjust the temperature on the Water Bath?

Follow  the  detailed  instructions  in  the  SET-UP 
INSTRUCTIONS,  "Item  #8  -  Changing  the  temperature 
on  your  Water  Bath".  This  will  explain  in  detail  how  to 
adjust the temperature.

 

--------------------------------------

How do I know what to adjust the temperature to?

Refer  to  splint  material  manufacturer  for  recommended 
temperature & duration of splint material used.

--------------------------------------

How long will it take for my Water Bath to reach the 
adjusted temperature?

Allow  60  to  75  minutes  for  the  appliance  to  reach   
temperature. Keep Lid closed to ensure quickest heat-up 
and to maintain set point temperature.

 

--------------------------------------

How do I open the lid more than 90

º

?

Make  sure  unit  is  unplugged  and  all  water  is  drained 
from  the  tank.  Take  an  adjustable  wrench  and 
screwdriver and remove the 2 screws and nuts from the 
lid stay on the underside of the lid. Remove the lid stay 
and  reinstall  the  screws  and  nuts  in  the  lid.  Follow  this 
procedure for both sides. 

Remove  the  hinge  stays  from  the  pins  located  in  the 
tank. Make sure to reinsert the screws. 

This will allow the lid to open past 90º and keep the lid in 
an upright position. Please keep these parts as they can 
be used to reassemble the lid stays to the lid if a partially 
open lid is again warranted.

FAQs

 

(Frequently Asked Questions)

 & TROUBLESHOOTING

Instructions for Operation and Care of Water Bath SP-1600-A-220

Summary of Contents for SP-1600-A-220

Page 1: ...SP 1600 A 220 WATER BATH 6900 124 000 August 2012...

Page 2: ...lacing removing splinting material AVOID skin contact with HOTwater at all times l DO NOTuse in the presence of flammable liquids or gases as these may present a fire and or explosion hazard l KEEPthe...

Page 3: ...ater Bath 3 Location of your Water Bath 3 Preparing to fill your Water Bath 3 Filling your Water Bath 3 Preparing to turn ON your Water Bath 3 Turning ON your Water Bath 3 Changing the temperature on...

Page 4: ...ire wound heating element The has an illuminated ON OFF power switch and a water resistant temperature controller conveniently located at the front of the appliance The temperature is easily adjustabl...

Page 5: ...eptacle A G F C I Ground Fault Circuit Interrupter receptacle will provide protection against electrical shock should water and electricity mix NOTE It is recommended that the electric receptacle used...

Page 6: ...on and work lid up and down until lid stays synchronize and allow lid to close USING YOUR WATER BATH EMPTYING YOUR WATER BATH CARE CLEANING OF YOUR WATER BATH Before attempting to clean the appliance...

Page 7: ...ly remove the bottom cover by removing the mounting screws The technician will troubleshoot and or replace the operating thermostat The heating element should be visually and electrically inspected to...

Page 8: ...rmostat is broken Splint material too hot The ON OFF switch is broken Replace the switch Turn on the circuit breaker Replace the thermostat Replace the heating element Circuit breaker is off Thermosta...

Page 9: ...and assistance in troubleshooting the appliance Be advised that altering the appliance s electrical wiring may affect the performance of the appliance and or render the appliance inoperable and may v...

Page 10: ...dustry CA 91744 0527 U S A Phone 800 782 7706 626 968 6681 Fax 626 855 4862 Web www whitehallmfg com 8 19 18 23 22 25 24 15 5 3 6 7 8 1 4 20 17 13 12 10 14 16 9 18 21 27 26 28 11 2 29 Instructions for...

Page 11: ...99 Bottom Cover Assembly 12 7000 415 000 Self Retaining J Nut 13 6502 100 000 8 32 x 3 8 Phil Pan Head Screw 14 6900 138 000 Pull Handle 15 6900 154 000 10 24 x 1 2 Pan Head Screw SST 16 0411 021 000...

Page 12: ...ot apply to materials which have been damaged by other causes such as mishandling or improper care or abnormal use The repair or replacement of the defective materials shall constitute the sole remedy...

Reviews: