Whitehall Para-Therapy PT-18-S Instructions For Operation And Care Download Page 2

WHITEHALL MANUFACTURING

 • P.O. BOX 3527 • City of Industry, CA 91744-0527 U.S.A

Phone (800) 782-7706 • (626) 968-6681 • Fax (626) 855-4862 • Web: www.whitehallmfg.com

a

            CAUTION

              DANGER

AVANT DE CONNECTER,

LIRE LES INSTRUCTIONS

!

!

BEFORE CONNECTING,

READ INSTRUCTIONS

Grounding reliability can only be achieved when the 
equipment is connected to an equivalent receptacle 
marked “Hospital Only” or  “Hospital Grade”.

Medical Equipment:
With respect to electric shock, fire and mechanical 
hazards only in accordance with UL  60601-1 and CSA 
C22.2 No. 601.1

La mise à terre ne fonctionne de facon efficace que si l’ 
équipment est connecté à une prise marquee “Hopital 
Seulement” ou “Categorie Hopital”.

Équipement Médical:
Concernant les risques de chocs électriques de feu et 
autres dangers mecaniques seulement conformément a  
UL  60601-1 et a CSA C22.2 No. 601.1

              WARNING

Risk of explosion if used in the presence of flammable 
anesthetics

l

REPLACE fuse as marked

l

Outer Surface is Hot

l

AVOID Prolonged skin contact.

l

AVOID resting objects on or against the tank.

Risque d’ explosion si utilisé en présence d’ 
anesthisques inflammables

l

REMPLACER les fusibles tel qu’ indiqué

l

La surface extérieure est chaude

l

ÉVITER tout contact prolongé avec la peau.

l

ÉVITER de deposer des objets sur le reservoir ou de             

les appuyercontre celui-ci.

l

RETIRER tout emballage, accessoire et feuillet se 

trouvant à l'intérieur de votre nouvel appareil avant de 
l'utiliser.

EFFECTUER LA MISE À LA TERRE de votre 
appareil DE FAÇON APPROPRIÉE. Brancher 
l'appareil dans une prise à disjoncteur différentiel de 
fuite à la terre (DDFT), ce qui offre une protection au 
cas où l'eau et la paraffine entreraient en contact 
avec l'électricité. Si l'appareil n'est pas bien mis à la 
terre, cela augmente les RISQUES DE DÉCHARGES 
ÉLECTRIQUES !

l

NE JAMAIS utiliser l'appareil en prenant un bain ou 
près de l'eau. RISQUE DE DÉCHARGES 
ÉLECTRIQUES !

l

NE JAMAIS utiliser l'appareil près de liquides ou de 
gaz combustibles. RISQUE D'EXPLOSION ! 

ÉVITER DE TOUCHER LE FOND DU RÉSERVOIR. 
Le fond peut être plus chaud que la paraffine en 
raison de l'emplacement des éléments chauffants.

l

l

FOR PROFESSIONAL USE ONLY

POUR PROFESSIONEL USE SUELEMENT

            ATTENTION

              DANGER

              AVERTISSMENT

l

PROPERLY GROUND your unit. Plug unit into a 
Ground Fault Circuit Interrupter (GFCI) receptacle for 
safety. Failure to properly ground unit can increase 
the risk of ELECTRICAL SHOCK HAZARD!

l

REMOVE all packaging, accessories and literature 
from inside of unit prior to using.

l

NEVER use unit while bathing or near water. 
ELECTRICAL SHOCK HAZARD!

l

NEVER use unit in the presence of combustible 
liquids or gases. EXPLOSION HAZARD!

l

AVOID TOUCHING BOTTOM OF TANK. Bottom can 
be hotter than paraffin due to heating element 
location.

Summary of Contents for Para-Therapy PT-18-S

Page 1: ...ra Therapy Revised April 2017 5900 100 000 INSTRUCTIONS FOR OPERATION AND CARE OF 800 782 7706 626 968 6681 www whitehallmfg com P O Box 3527 City of Industry CA 91744 0527 U S A Division of Acorn Eng...

Page 2: ...eservoir ou de les appuyercontre celui ci l RETIRER tout emballage accessoire et feuillet se trouvant l int rieur de votre nouvel appareil avant de l utiliser EFFECTUER LA MISE LA TERRE de votre appar...

Page 3: ...ams and Specifications Model PT 6 S Page 6 Model PT 18 S Page 8 Repair Parts Model PT 6 S Page 6 Model PT 18 S Page 8 Exploded View Model PT 6 S Page 7 Model PT 18 S Page 9 Troubleshooting Guide Page...

Page 4: ...ns 13 L x 7 W x 7 3 4 D 330 L x 180 W x 195 D Inner Tank Dimensions 11 3 4 L x 5 3 4 W x 5 3 4 D 300 L x 145 W x 145 D Tank Capacity 6 lbs 2 72 kgs Electrical Requirements Inlet Voltage 120 VAC Freque...

Page 5: ...een pound unit will take 10 to 12 hours to melt all the paraffin completely Once paraffin is completely melted your unit is ready to use Operating Tips When do I need to add Paraffin Add paraffin when...

Page 6: ...s of the body which cannot be dipped into the tank a soft brush may be used to apply multiple layers of paraffin to area as required for treatment 5 For maximum benefit after the paraffin has been app...

Page 7: ...towel Carefully lift unit up an inch or two and then bump it down on the surface to break free the solid paraffin block Block should drop on to lid Lift away unit leaving the paraffin block on top of...

Page 8: ...113 001 Euro Outer Housing 13 6505 540 000 Indicator Light 250 VAC 14 5902 113 199 Bottom Cover 15 6504 604 000 Snap In Bumper Foot 16 0331 014 000 0 203 ID x 0 734 OD Stainless Steel Washer 17 0116...

Page 9: ...855 4862 Web www whitehallmfg com Instructions for Operation and Care of Paraffin Baths FUSE 1 FUSE 2 GND INLET L N THERMO INDICATOR HEATING ELEMENTS LIGHT STAT THERMAL CUTOUT POWER Electrical Specifi...

Page 10: ...13 199 Bottom Cover 14 6505 540 000 Amber Indicator Light 125 V 15 0116 010 000 10 32 x 1 2 Phillips Truss Head Screw 16 0302 004 000 10 32 Hex Nut 17 6502 560 000 10 Stainless Steel Lockwasher 6504 6...

Page 11: ...68 6681 Fax 626 855 4862 Web www whitehallmfg com Instructions for Operation and Care of Paraffin Baths 1 2 3 6 7 23 24 5 3 4 2 9 8 3 10 11 12 13 14 15 16 17 18 19 20 12 25 FUSE 1 FUSE 2 GND INLET L N...

Page 12: ...MAL CUTOUT BURNT OR FAULTY REPLACE THERMOSTAT THERMOSTAT IS BROKEN HEATING ELEMENT IS REPLACE HEATING ELEMENTS BURNT OR FAULTY PARAFFIN WAX IS MELTING INDICATOR LIGHT REPLACE INDICATOR LIGHT INDICATOR...

Page 13: ...stitute the sole remedy of the Buyer and the sole remedy of Whitehall under this warranty Whitehall shall not be liable under any circumstances for incidental consequential or direct charges caused by...