Whip Mix 05350 Operating Instructions Download Page 2

installation

Each Whip Mix Water Bath is packaged as follows:
 

1 – Water Bath unit

 

1 – Removable pan

Fill the removable pan with water to the water level indicator line.
Place removable pan in the Water Bath Unit by positioning the pan into the opening and rotating 
the two clips over the lip of the pan.

Wrap gauze around the Lift Tray (see Accessories) to prevent materials such as impression  
compound from sticking to the tray and removable pan surfaces. 

Plug the unit into a grounded receptacle having the same electrical characteristics as shown  
on the Water Bath data plate.

enviRonmental conditions

This Water Bath has been designed to be safe at least under the following  
environmental conditions:

•  Indoor use.
•  Altitude up to 6,500 feet.
•  Temperature 41°F to 104°F.
•  Maximum relative humidity 80% for temperatures up to 88°F decreasing  

linearly to 50% relative humidity at 104°F.

•  Mains supply voltage fluctuations not to exceed plus or minus 10%  

of the normal voltage.

•  Transient overvoltages according to InsTAllATIon CATegory II.
•  PolluTIon Degree 2 in accordance with IeC 664.

opeRation

Turn the unit on by pushing the green on/off switch. 

Note:

 The unit will not turn on until the  

removable pan is correctly positioned in the Water Bath Unit.

set unit to desired temperature scale (Celsius or Fahrenheit) by pressing the °C/°F button.  

Note:

 The Heating Indicator light will glow continuously until the desired water temperature  

is reached.

Press the C/s button to illuminate the "s" light. Then press the blue arrow buttons  
  

for increased heat or  

to decrease. 

To verify the actual water temperature, press the C/s button again to illuminate the "C" light. This 
will show the actual temperature.

After clinical procedures are completed or at the end of the day, turn the unit off by pushing the 
green on/off switch. Remove the lift tray from the removable pan if being used and discard gauze. 

gently lift removable pan from the Water Bath unit and discard water.

Autoclave lift tray and removable pan. 

tecHnical speciFications

Voltage: 

115 V/60 Hz

Amperage: 

1.5 A

Wattage: 

175 W

Dimensions: 

7.5" (19 cm) Wide x 5" (12.7 cm) High x 8.75" (22.2 cm) Deep

Capacity: 

6.5" (16.5 cm) Wide x 4.25" (10.8 cm) High x 2.75" (7.0 cm) Deep

Temperature: 

90°F to 160°F (32°C to 71°C) Variable

Weight: 

5 lbs. (2.3 kg)

maintenance

Unplug the unit before performing any maintenance or repair.

Keep the gasket on top of unit free of debris so that any spilled water does not enter the area 
containing the electrical components.

return unit to Whip Mix Corporation for service on all internal components.

caUtion

•  Water level within the basin must be maintained to cover the thermometer.

•  This appliance must be turned off at the end of each day. Failure to do so may cause  

the water to evaporate and the unit to overheat.

•  When cleaning this appliance, care must be taken to avoid damage to the thermostat. 

Damage to this component may cause the unit to overheat.

•  WArnIng:  To prevent electrical shock, Do noT reMoVe ground prong on the plug.  

Use only in a properly grounded outlet.

• 

WARNING:  Allowing this appliance to operate without water or with damage  
to the thermostat bulb or tube can cause the unit to overheat, causing damage 
to the unit, shock or fire hazard

.

•  WArnIng:  Do not immerse water bath in water. unplug the unit before performing any 

maintenance or repair.

17125_Water_Bath_Manual.indd   2

4/26/11   10:33 AM