![WHALETEQ MECG 2.0 User Manual Download Page 30](http://html.mh-extra.com/html/whaleteq/mecg-2-0/mecg-2-0_user-manual_967657030.webp)
MECG 2.0 Test System | User Manual
28
Figure 19: View ECG waveforms
8
Trouble Shooting
Table 6: Trouble Shooting
Problem
Resolution
USB module (test unit)
not recognized (USB
driver is installed
correctly)
Recognition of USB devices needs to be done in
order:
1) Close WhaleTeq software if open
2) Disconnect the USB module for ~2s
3) Reconnect the USB module
4) Wait for the recognition sound
5) Start WhaleTeq software
USB streaming is
interrupted
(occasional)
USB Module with firmware before Aug 2011
The system automatically detects streaming delays,
attempts to move the system to “Off” mode and
provides the user with a warning. To resume
operation, restart the function that was being
previously used. In some cases, it may be
necessary to restart software / USB module.
USB Module with firmware from Aug 2011
The system automatically detects streaming
delays, but continues to stream data and simply