![Wet Labs ac-s User Manual Download Page 27](http://html1.mh-extra.com/html/wet-labs/ac-s/ac-s_user-manual_966429027.webp)
ac-s User’s Guide (ac-s)
Revision F
9 Feb. 2006
19
5.
Specifications
Mechanical Parameters
Size
31 x 4.1 in diameter (79 x 10.4 cm diameter)
Weight (in air)
13 lbs (5.9 kg)
(in water)
2.5 lbs (1.1 kg)
Electrical Parameters
Input 10–35
VDC
Current draw
0.83 Amps @ 12 V nominal
Serial output
RS-232, -422, or -485
Connector MCBH6M
Sample rate
4 scans/sec., nominal
Environmental Parameters
Temperature range
0–30 deg C
Rated depth
500 m
Pressure sensor
optional
Optical Parameters
Spectral range
400–730 nm
Band pass
15 nm/channel
Pathlength
25 cm (10 cm available)
Beam cross-section
8 mm dia. (nominal)
Linearity
≥
99% R
2
Output wavelengths
80–90
Resolution 4
nm
Precision (450–750nm)
+/-0.001 m
-1
typ., 0.003 max @ 4 Hz
+/-0.0005 m
-1
typ., 0.0015 max @ 1 Hz
Precision (400–449 nm) +/-0.005 m
-1
typ., 0.0012 max @ 4 Hz
+/-0.003 m
-1
typ., 0.006 max @ 1 Hz
Accuracy
+/- 0.01 m
-1
Dynamic range
0.001–10 m
-1
5.1
Connectors
The ac-s has two bulkhead connectors mounted on the top flange. The power and data
connector has six pins and connects to the sea or test cable. It provides input power to the
instrument and data output in RS-232 and RS-485 formats. A second connector with
three sockets provides power to the pump (Figure 3). Pin or socket numbers are molded
into the rubber bodies of the plugs (Figure 4 and Figure 5).