7.  Troubleshooting

Please contact our technical support team on 1800 358 101. 

Error

Issue

Solution

Droplets left 

inside the 

tip

Unsuitable tip, non-

uniform wetting of  

the plastic

Use a new tip.

Leakage 

or pipetted 

volume is 

too small

Tip cone is scratched  

or damaged

Change the tip cone.

Organic solvent as liquid

Aspirate and discard the organic 

solvent several times before actually 

pipetting with the same tip.

Tip incorrectly attached

Attach the tip firmly.

Unsuitable tip

Use a new tip.

Foreign particles between 

the tip and the tip cone

Clean the tip cone.

Insufficient amount of 

grease on the piston

Clean and grease the piston.

Inaccuracies

Incorrect operation

Follow instructions carefully,

Calibration altered

Recalibrate according to the 

instructions. 

Unsuitable for the 

particular liquid pipetting 

technique

Use the correct pipetting technique.

Pipette damaged

Contact our technical support team  

on 1800 358 101.

Push button 

jammed 

or moves 

erratically 

Piston contaminated

Clean and grease the piston.

Penetration of solvent 

vapours

Tip ejector 

is jammed 

or moves 

erratically

Tip cone is contaminated 

from the outside

Remove the ejector collar and clean 

the outer surface of the tip cone with 

ethanol.

Volume 

setting 

isn’t ‘click-

stopping’ 

properly

Click stop mechanism is 

damaged

Contact our technical support team  

on 1800 358 101.

Push button 

doesn’t turn 

for volume 

setting

Use of excessive force 

beyond the range of the 

pipette

Contact our technical support team  

on 1800 358 101.

Summary of Contents for 665-251

Page 1: ...1 PRODUCT MANUAL Precision Single Channel Pipettes Codes 665 251 665 252 665 254 665 255 665 256 665 257 665 258...

Page 2: ...resistant material allowing them to be safely autoclaved multiple times without a loss of quality With a fully autoclavable body the transfer of contaminants during pipetting is reduced The pipettes...

Page 3: ...drops for colour coding Bag of sample tips Instruction manual Calibration report 2 Technical Specifications Code Variable Volume Increments Test Volume Inaccuracy Imprecision 665 251 0 5 10 l 0 1 l 1...

Page 4: ...es that are not supplied with the unit Is operated with accessories that are not supplied or recommended by the manufacturer Is operated improperly or contrary to the manufacturer s specifications 4 M...

Page 5: ...ssembly parts Number Name 1 Push button 2 Top closure 3 Spring stopper 4 Primary spring support 5 Primary spring 6 Main body 7 Counter Number Name 8 Washer 9 Shaft 10 Bottom closure snap fitting 11 Ti...

Page 6: ...er Name 1 Piston 2 Piston spring 3 Piston support 4 Tip cone tip 5 O ring 6 Tip cone 7 Ejector collar Number Name 8 Secondary spring 100 1000 l 9 Secondary spring support 10 Washer 500 5000 l 11 Washe...

Page 7: ...ia New Zealand United States Canada Calibration disassembling tool The calibration disassembling tool is supplied for maintenance and calibration of the pipette Pipette identification The pipettes are...

Page 8: ...tool between the tip ejector and the tip ejector collar to release the locking mechanism 3 Carefully release the tip ejector and remove the ejector collar 4 Place the wrench end of the opening tool o...

Page 9: ...d dispense the liquid To set the volume 2 Top closure Used to adjust the calibration with the help of the calibration tool 3 Drop The drop is colour coded for easy identification 4 Tip ejector The tip...

Page 10: ...consistency Allow the liquid tip and pipette to equilibrate to the ambient temperature before use When not in use place the pipette on a bench to avoid transferring body heat resulting in incorrect di...

Page 11: ...and United States Canada Loading and ejecting tips 1 Ensure that the tip cone is clean 2 Press the tip firmly onto the tip cone to ensure an airtight seal 3 The pipettes are equipped with a tip ejecto...

Page 12: ...will empty the tip completely and ensure accurate pipetting 5 Release the push button to the ready position Reverse pipetting Reverse pipetting is recommended for viscous solutions with a tendency to...

Page 13: ...For variable volume pipettes at least three volumes shall be tested nominal volume 50 of nominal volume and 10 of nominal volume 1 Select the test volume Do not change the setting during the test cyc...

Page 14: ...etween the tip and the tip cone Clean the tip cone Insufficient amount of grease on the piston Clean and grease the piston Inaccuracies Incorrect operation Follow instructions carefully Calibration al...

Page 15: ...lab com au 24 hours 7 days a week convenience 15 PRODUCT MANUAL 1800 358 101 sales au westlab com www westlab com au 64 9 553 3677 sales nz westlab com www westlab co nz 646 568 5391 sales us westlab...

Page 16: ...Westlab in every discovery...

Reviews: