Digital Display Overhead Stirrer 40L Set
Code 665-110
Page 1: ...1 PRODUCT MANUAL Product Manual Digital Display Overhead Stirrer 40L Set Code 665 110 ...
Page 2: ...s in liquids the stirring of volatile solutions and other homogeneous mixing 1 3 Features Brushless DC motor for quiet operation and maintenance free long life Microprocessor controlled stirrer with variable speed and time setting Timer setting ranges from one minute to 99 hours 59 minutes and continuous mode Clockwise CW or counter clockwise CCW motion Adjustable speed in steps of 1 RPM Corrosion...
Page 3: ...es au westlab com www westlab com au 64 9 553 3677 sales nz westlab com www westlab co nz 646 568 5391 sales us westlab com www westlab com 1 877 822 1455 sales ca westlab com www westlab com Australia New Zealand United States Canada ...
Page 4: ...rring Quantity H2O 40L Viscosity max 30 000 mPas Torque at Chuck 100 RPM 40 Ncm Speed Range 50 RPM to 1 000 RPM Speed Setting Accuracy 1 RPM Timer Range 1 minute to 99 hours 59 minutes and continuous mode Display Digital LED Reversible Direction of Rotation Yes Power Intermittent Operation Yes it resumes from left parameters upon resuming power Chuck Range 0 7mm to 10mm Protection Class IP 54 DC V...
Page 5: ... on the unit during the motion of the impeller Use a large container beaker or flask to avoid spillage Place the device on a flat stable clean non slippery and fireproof surface The base plate must be kept clean and intact Do not try to mix liquids more than the recommended volume of liquid H2O at room temperature Wear protective equipment in accordance with the hazardous category of the media to ...
Page 6: ...hat are not supplied with the unit Is operated with accessories that are not supplied or recommended by the manufacturer Is operated improperly or contrary to the manufacturer s specifications If the printed circuit board is modified by a third party 3 2 Maintenance Wear protective gloves when cleaning the unit Always clean the impeller after every use for improved operation Regularly check that t...
Page 7: ...from the box 2 Place the unit gently onto a firm and levelled surface 3 Unpack every individual component including the main unit power adaptor impeller stand Bosshead 4 Assemble the components as instructed in this manual for optimal performance ATTENTION The manual should be kept with the device and all the packaging should be kept in safe storage for at least two years for warranty purposes Ins...
Page 8: ... button By default the motion will be set to the clockwise direction 3 Time Press the Time button to set the value between one minute and 99 hours 59 minutes or continuous mode 4 Increment Press the Increment button once to increase the value by an increment of one Press and hold the button to increase the value faster 5 Decrement Press the Decrement button once to decrease the value by an increme...
Page 9: ...es on the display Power intermittent mode The power intermittent mode allows the user to set operations to auto resume from the remaining time upon the restorage of power If there is a power failure or mechanical fault it allows the user to start where they had been interrupted during the failure By default this mode is switched OFF Press the Power Inter button once to activate the mode A LED ligh...
Page 10: ...the load on the motor and restart the overhead stirrer Error OT Over Temperature The motor temperature is too high 1 Switch off the device completely 2 Allow the unit to cool down fully 3 Restart Error 52 No motor feedback Contact a service technician at Westlab Pty Ltd Vibrates too much and the unit emits excessive noise during operation A bent or damaged impeller Check the impeller A bent or dam...
Page 11: ...11 PRODUCT MANUAL This page has been left blank intentionally ...