1
PRODUCT MANUAL
Product Manual
Westlab Digital Thermostatic Water Baths
Codes 663-890, 663-891, 663-892, 663-894, 663-895, 663-896
Page 1: ...1 PRODUCT MANUAL Product Manual Westlab Digital Thermostatic Water Baths Codes 663 890 663 891 663 892 663 894 663 895 663 896...
Page 2: ...a removable sturdy low height stainless steel perforated shelf for even heat distribution The shelf provides a deeper volume of water for standing containers and accessory test tube racks in the bath...
Page 3: ...cation 663 890 663 891 663 892 Tank Capacity 4L 14L 22L Temperature Range Ambient 5 C to 99 C Temperature Control Digital LED Display Actual or Set Temperature Resolution of 1 C Water Level Range 55mm...
Page 4: ...ge to that specified on the rating label positioned on the rear of the unit 9 Do not block ventilation slots during use and always follow installation instructions 10 Ensure substances being heated pr...
Page 5: ...emble rust These are usually small ferrous particles which have oxidised Using a stainless steel cleanser such as Bar Keepers Friend will remove these marks as well as mineral deposits Exterior Anti B...
Page 6: ...own Arrow To Decrease Set Point Value U To Access Temperature Readings 4 Press the P key Display will alternate between SP set point and the set value 5 Use the UP and DOWN ARROW keys to adjust the va...