3. Operational Parameters 

 

The Operational Parameters or Duty Cycle for the 

operation is set to avoid damage to the internal coil or the Output Module, and must 
be observed. 

 

AC Operation:

  It is recommended that the operator does not keep the Yoke on for 

more than 5 minutes at a time, as the Yoke housing may get to warm to hold.  
However, the basic design of any Yoke inherently produces heat. 
Typical operation is 5 - 15 seconds on, while applying inspection media, followed by 
5 - 15 seconds off repositioning the yoke to the target area. 

 

If the Yoke is used for prolonged periods of time such as 60 to 90 minutes of 
continuous cycling, as outline above, the Yoke will get hot.  If the WC-6

X

2 is used in 

this manner the operator must provide time for a sufficient cooling period, or 
components in the Electronic Control Module may fail.

 

 

Do not hold the Push Button Switch on when plugging the Yoke into power.

 

 

4. Field Characteristics

 

AC Field –

 AC Magnetic Fields are sensitive to surface and near surface defects 

due to the ‘Skin Effect’ as the field travels from one Pole Piece to another.  The WC-
6

X

2 is designed to produce a very strong AC field, to aid the operator while 

inspecting though nonmagnetic coatings, or using extra scrutiny for inspecting 
critical components.  The Inspection Media (Dry Powder or Wet Method Particles) 
has a tendency to migrate toward interruptions (or defects) in the field.  The direction 
and intensity of an AC Field, by it’s nature, alternates causing high particle mobility, 
so defects tend to be revealed immediately. 

Demagnetization – 

Small Parts may be demagnetized by positioning the contact 

surfaces of the Pole Pieces together, activating an AC Field and pass the part 
through the opening formed between the Legs and Yoke Housing.  Larger Work 
Pieces can be demagnetized by placing the Yoke on the surface, in a similar 
manner used during inspection, activating an AC Field and pull the Yoke off the 

WC-6X2 Operation Manual 

  End Cap

 (Name 

Plate and Fastener 
     not shown)

Cord Protector/ 
  Strain Relief 

Ground Lug 

  (Not used) 

Cross Bar

 

  Junction Tube 

  Module and  
Insulating Tape 

Rear Leg 

Power Lines 

White – Positive 
Black - Negative 

Summary of Contents for WC-6X2

Page 1: ...Operating Instructions WC 6X2 High Output Compact AC Yoke WC 6X2 Operation Manual 1 E s t a b l i s h e d 1 9 6 5 ...

Page 2: ...ctronic controls in the th itch on when plugging the Yoke into power Do not hold the Push Button Sw 2 End Cap Rotation The End Cap may be rotated to reposition the Cord Protector Strain Relief for operator convenience within the Bottom 120º or Top 120º of the Junction tube Carefully remove the End Cap followed by the internal Cross Bar and reposition the power wires Replace the cross bar ensuring ...

Page 3: ...ve to surface and near surface defects due to the Skin Effect as the field travels from one Pole Piece to another The WC 6X2 is designed to produce a very strong AC field to aid the operator while inspecting though nonmagnetic coatings or using extra scrutiny for inspecting critical components The Inspection Media Dry Powder or Wet Method Particles has a tendency to migrate toward interruptions or...

Page 4: ...he Pole Pieces farther apart If work piece configuration does not permit extending Pole Pieces reduce the contact area of the Pole Pieces on the work piece Follow the Operational Parameters outlined in these instructions 6 Maintenance After extended use the Yoke should be cleaned with a mild soap solution The unit should be visually inspected for any damage that could cause harm to the operator or...

Page 5: ...anty is void if the unit has been modified in any way or if it has been repaired by an unauthorized agency The end user agrees that any equipment s disposition when returned for warranty work is at the full discretion of Western Instruments as to whether a claim is under warranty or due to misuse Western Instruments warranty shall overlook normal wear however does not include operation outside the...

Page 6: ...overs Junction Tube Housing Hand Grip Protects Internal Coil Switch Cavity Hidden from View Name Plate Model Serial Number Volts Amps Frequency Cord Protector Strain Relief Power Cord AWG18 3 SOOW Rear Leg Assembly Pole Pieces Front Foot Upper Leg Front Leg Assembly Pole Pieces Upper Leg Rear Foot E s t a b l i s h e d 1 9 6 5 ...

Reviews: