5.  Operation:

 

Position the Pole Pieces (Feet) on the work piece.  The area between the pole pieces 
is your target area, which also extends laterally out, approximately 1.5” (38mm), from 
either edge of the pole pieces.  The Field will expose defects that are transverse to 
the centerline between the Pole Pieces.  The Pole Pieces should be positioned so 
that as much of their contact surfaces as possible, are on the work piece.  The Yoke 
is then energized, by pressing Push Button Switch, and Magnetic particles are 
applied.  Dry Method Particles are dusted between the Pole Pieces and over the 
target area, while Wet Method Particles are sprayed in a similar manner. 

 

The Target Area is then inspected visually for a collection of Particles around  
defects.  A Black Light is used to aid visual inspection when Fluorescent Particles are 
used.  Indications found with Dry Powder will tend to form immediately, and will take 
slightly longer with Wet Method Particles.  If the typical direction of defects is not 
known, rotate the Yoke through 90º and repeat the inspection of the target area. 

 

The WC-6 produces a standard amount of Field Blow as other AC Yokes.  Field Blow 
is a collection of Inspection Media between the Pole Pieces, transverse to the 
centerline between the Pole Pieces, and may case a masking of indications.  Field 
Blow can be minimized by extending the Pole Pieces farther apart, If work piece 
configuration does not permit extending Pole Pieces, reduce the contact area of the 
Pole Pieces on the work piece.  Follow the Operational Parameters outlined in these 
instructions. 

 

6. Maintenance: 

After extended use the Yoke should be cleaned with a mild soap solution and 

thoroughly dried.  The unit should be visually inspected for any damage that could 
cause harm to the operator, or the material being inspected.  Special attention should 
be paid to the Push Button Switch Cover, to ensure it is fully adhered to the body of 
the Yoke.  Furthermore, the Power Plug, Power Cord, and the End Cap/Cord 
Protector should be in a good state of repair.  Before performing maintenance, 
cleaning, or repositioning the End Cap, the Yoke should be disconnected from any 
power source, with safe industrial practices employed.  Any potential problems to 
these assemblies must be reported to the Distributor or Western Instruments for 
instructions on corrective action. 

 

Whether industrial specifications are being observed or not, the Yoke should be 
tested periodically, using a certified Pull Test Bar such as the W-Series W-PT

®

, to 

ensure it continues to lift the specified amount of weight.  If the unit fails such a test, 
first inspect the Pole Pieces to ensure they fully contact the test weight.  If the unit 
continues to fail, contact the Distributor or Western Instruments for instructions on 
corrective action.  

 

Wiring:

 

W-Series 230 Volt Models, are designated by a “K” placed after the Serial Number 
and the Model number (e.g. WC-6K), are shipped without an AC Power Plug as there 
is no international standardization.  When installing an AC Power Plug onto the AWG 
18-3 Power Cord, the following is the identity of the 3 Color Coded Conductors;  

WC-6 Operating Instructions

 

4

Summary of Contents for WC-6

Page 1: ...59 6720 St Albert Alberta Fax 780 459 7837 T8N 1M8 E mail info westerninstruments com Web www westerninstruments com Operating InstructionsDecember 2007 E s t a b l i s h e d 1 9 6 5 WC 6 Compact AC Yoke WC 6 Operating Instructions ...

Page 2: ... 115V 60Hz and Red are 230V 50Hz Do not hold the Push Button Switch on when plugging the Yoke into power 2 End Cap Rotation The End Cap may be rotated to reposition the Cord Protector Strain Relief for operator convenience within the Bottom 120º or Top 120º of the Junction tube Carefully remove the End Cap followed by the internal Cross Bar and reposition the power wires Replace the mounting bar e...

Page 3: ...nic Control Module may fail Do not hold the Push Button Switch on when plugging the Yoke into power 4 Field Characteristics AC Field AC Magnetic Fields are sensitive to surface and near surface defects due to the Skin Effect as the magnetic field travels from one Pole Piece to another The Inspection Media Dry Powder or Wet Method Particles has a tendency to migrate toward interruptions or defects ...

Page 4: ...of the Pole Pieces on the work piece Follow the Operational Parameters outlined in these instructions 6 Maintenance After extended use the Yoke should be cleaned with a mild soap solution and thoroughly dried The unit should be visually inspected for any damage that could cause harm to the operator or the material being inspected Special attention should be paid to the Push Button Switch Cover to ...

Page 5: ...nstruments warranty shall overlook normal wear however does not include operation outside the environmental specification of the product All warranty work is FOB Western Instruments and any returned units shall include a written description by the end user of the fault Western Instruments makes no other warranty either expressed or implied with respect to this product Western Instruments specifica...

Reviews: