9 WS-CH4-TLD Methane detector
PAG
E
67
w
w
w
.w
es
tsy
stems
.co
m
W
EST
Syst
em
s
In order to guarantee the measuring of CH4 concentration in a wide range (0-
100% vol), the sensor needs to apply different settings on the laser diode.This
adaptations produce a gap of a few seconds during which a valid concentration
value is not available.
9.2
Computing flux
The sensor presents a first change of scale around 140-150 ppm and a second
change around 2300-2400 ppm. During a change of scale, the value remains
constant for about 10 seconds.