Appendix
76/82
7
Appendix
7.1
wellsystem MEDWAVE TOUCH: EMC
information
7.1.1
Table 201: Guidelines and manufacturer's
declaration – Electromagnetic emissions
The wellsystem MEDWAVE TOUCH massage device has been designed for
use in the electromagnetic environment specified below. The device operator
must ensure that it is used such an environment.
Emission measurements
Conformity
Electromagnetic environment – Guidelines
HF emissions as per CISPR 11
Group 1
wellsystem MEDWAVE TOUCH uses HF power solely
for internal functions. Therefore the HF emissions are
very low and it is highly unlikely that it will cause any
interference to adjacent electronic equipment.
HF emissions as per CISPR 11
Class B
wellsystem MEDWAVE TOUCH has been designed for
use in environments of professional healthcare facilities
that are connected to a public mains supply that also
supplies buildings which are used for residential pur-
poses.
Harmonics
IEC 61000-3-2
Class A
Voltage fluctuations / Flicker
IEC 61000-3-11
Fulfilled
The obligations listed in Sub-section 6.2.2 of EN 61000-
3-11:2000 must be taken into consideration. In particu-
lar, the 0.14 ohm line impedance must be guaranteed
for pulse massaging.