22
IMPORTANT
• The code card found in the Wellion LEONARDO Cholesterol Test Strip package is for use
with that particular package only. Different packages of Wellion LEONARDO Cholesterol
Test Strips and other brands of test strips will produce inaccurate results.
• Re-coding needs to be done when a new package of test strips is opened for use.
• The code number will be memorized by the meter even if the code card is removed.
NOTE
• If the code number displayed on the screen does not match the code number printed on
the test strip foil pouch or package, it will create inaccurate test results.
• If error messages (E 1, E 2, etc.) appear, see “Solving Problems”.
Begin testing by inserting a test strip, or turn off the meter by pressing and holding “OK” until the
meter display screen shows “OFF”.
into the meter. The meter will turn on automatically.
Step 6: A 3-digit code number will be displayed on the meter display screen. Verify the
code number on the screen with the number on the code card and the test strip foil
pouch. All three sets of numbers should be the same.