32
Power Source:
Four 1.5V DC “AA” alkaline batteries
(not included)
or 6V DC AC adapter
Measurement Method: Oscillometric
Measurement Range:
Pressure: 40~280 millimeters mercury (mmHg)
Pulse: 40~199 beats/minute
Accuracy:
Pressure: ±3 mmHg
Pulse: ±5% of reading
Pressure Sensor:
Semi-conductor
Inflation:
Pump driven
Deflation:
Automatic Pressure release valve
Memory Capacity:
60 memories for each User (120 total)
Auto-shut-off:
1 minute after last button operation
Operation
Environment:
Temperature: 41°F~104°F (5°C ~40°C)
Humidity: 15% ~ 93% RH
Storage/Transportation
Environment:
Temperature: -13°F~158°F (-25°C~70°C)
Humidity: Less Than 93% RH
Weight:
0.69 lbs (310 g) (without batteries)
Arm Circumference:
Standard size cuff: 9”–13” (23-33 cm)
Large size cuff:13”–17” (33-43 cm)
Dimensions:
4.21”(L) x 5.55”(W) x 2.48”(H)
107mm(L) x 141mm(W) x 63mm(H)
Accessories:
6V DC AC adapter, 2 arm cuffs with tubes
Internally Powered equipment
Type BF Equipment
IPx0-No special protection against the
ingress of water
Not suitable for use in presence of flammable
anesthetic mixture with air, oxygen or
nitrous oxide.
Continuous operation with short-time loading.
Follow instructions for use.
SPECIFICATIONS
Note: These specifications are subject to change without notice.
IB-WGNBPA-740_A--02.indd 32
2013/8/7 下午8:05