WELCH-ILMVAC DRYFAST 2014 Owner'S Manual Download Page 9

9

 

The use of a cold trap when pumping heavy vapor loads will eliminate the need for a liquid 
trap attached to the exhaust port. The cold trap is installed between the pump and the vacuum 
chamber. The water or organic vapors evolved from the vacuum process will upon entering 
the cold trap will come in contact with the surfaces of the trap and condense. Commonly used 
refrigerants are liquid nitrogen or dry ice with alcohol slurry. Dry ice provides suffi cient cooling 
to freeze out most heavy water vapor loads. A variety of cold traps are available from Welch. 
Please call our customer service department. For additional information at (847) 676-8800.

2.19  The Care of a Liquid Trap
 

A liquid trap needs no refrigerant. The key maintenance issue when pumping high vapor loads 
is to regularly drain the trap of liquid ejected from the dry vacuum pump.

2.20  The Care of a Cold Trap
 

When using a cold trap the refrigerant should be maintained at a high level in the fl ask to keep 
the trap at a uniformly low temperature. If the trap is rewarmed it may allow re-evaporation of 
the condensate. The refrigerant add tube on the liquid nitrogen trap should not be obstructed 
as the refrigerant boil-off can produce dangerously high pressures. If the trap becomes 
saturated it should be disconnected from the system, drained and cleaned. An increase in 
pressure in the vacuum system will normally indicate that the trap has become saturated. 
To clean the trap, remove the trap from the system and allow the trap to warm up and rinse 
off the condensate with a suitable solvent in a fume hood. Thoroughly clean and dry the trap 
before reinstalling into the system. If a liquid nitrogen trap is used, the refrigerant add tube 
on the liquid nitrogen trap should not be obstructed as the refrigerant boil-off can produce 
dangerously high pressures.

Dry Vacuum Pump

Dry Vacuum Pump

Liquid 

Trap

Cold Trap

Rotary Evaporator

Rotary Evaporator

To Fume 
Hood

Summary of Contents for DRYFAST 2014

Page 1: ...tify intake and exhaust before using the pump See Section 2 15 CAUTION Do not pump liquids with the pump Pumping liquids will cause the pump to stop working WELCH ILMVAC Welch Ilmvac 5621 W Howard Str...

Page 2: ...ons are found where failure to observe the instruction could result in damage to the equipment associated equipment and process These units conform to the SI International system of units of measureme...

Page 3: ...he care of a cold trap 9 Section 03 Operation 3 10 Starting a Welch DRY Vacuum Pump 10 3 11 Cleanliness 10 3 12 Leak detection 10 3 13 Operating pressure range 10 3 14 The effects of unwanted vapor 10...

Page 4: ...tector resets Always disconnect power source before servicing 1 15 Never drop or insert fingers or any other object into any openings 1 16 Do not operate this product where oxygen is being administere...

Page 5: ...uid Unplug immediately 1 33 Never operate this product outdoors in the rain or in a wet area 1 40 Danger To Reduce Risk of Explosion or Fire 1 41 Do not use this pump in or near explosive atmospheres...

Page 6: ...he carrier If you are required to communicate with your dealer or Welch Vacuum be sure to include your order numbers for quick identification Do not return the pump to the factory without obtaining re...

Page 7: ...p a vent line can be attached which allows gases and vapors pumped through the pump to be piped from the work area into a hood Vent lines will muffle noise coming from the vacuum pump WARNING Never bl...

Page 8: ...a lubricating and sealing fluid for pulling a vacuum When pumping gases or low vapor loads a trap is not necessary unless you want to capture vapors evolved from the process When a heavy load of wate...

Page 9: ...Cold Trap When using a cold trap the refrigerant should be maintained at a high level in the flask to keep the trap at a uniformly low temperature If the trap is rewarmed it may allow re evaporation...

Page 10: ...y slightly pressuring the system and painting the suspected area with a thick soap solution Escaping air will produce soap bubbles 3 13 Operating Pressure Range Vacuum pumps are designed to be run fro...

Page 11: ...e cover plate and head from the plastic head base 2 Make sure to wear gloves before removing the diaphragm Grab the diaphragm with both hands and turn counterclockwise to loosen from connecting rod 3...

Page 12: ...for Models 2037 and 2047 to a standard water aspirator See the table below to cross reference the correct Welch Dry Vacuum pump with your water aspirator 5 20 Cross Reference Welch Dry Vacuum Pump wit...

Page 13: ...01 2014B 01 with North American 115V Plug Catalog Number Wired for 230V 50Hz 1 Phase 2032C 02 2042C 02 2034C 02 2044C 02 2037C 02 2047C 02 2014C 02 with European Schuko Plug Catalog Number Wired for...

Page 14: ...er short will be repaired under this warranty rather than replaced The warranty is effective for one year from the date of original purchase when 1 The warranty card has been completed and returned 2...

Page 15: ...25 6 26 8 59 7 25 11 64 3 11 6 26 6 39 7 32 8 66 5 30 14 89 3 18 6 32 MODELS 2032 2034 2037 2042 2044 2047 MODEL 2014 HOSE BARB FOR 3 8 HOSE I D HOSE BARB FOR 1 4 HOSE I D AVM CONTROL HOSE BARB FOR 3...

Page 16: ...16 8 20 Exploded View and Kit for Model s 2034 2037 2044 This Kit is a one Head Service Kit...

Page 17: ...17 8 30 Exploded View and Kit for Model s 2014 2032 2042 2047 This Kit is a one Head Service Kit...

Page 18: ...L 60714 Phone 847 676 8800 Technical Support Fax 847 677 8606 E Mail gdwelchvacuum gardnerdenver com Web Page www welchvacuum com Copyright 1997 2007 Gardner Denver Welch Vacuum Technology Inc Welch i...

Reviews: