Welch Allyn Inc.
Aneroid Sphygmomanometers Service Manual
Revision A
Welch Allyn Inc.
Page 20 of 40
Section 5: Economy Aneroid Gauge
5.1 Opening/Removing and Replacing Parts of the
Economy Gauge*
1. Remove the two screws from the back of the case using a Phillips
head screwdriver. Aside the screws and the clip.
2. Remove the Economy cover an d crystal and aside.
3. Remove the pointer using the pointer removal tool (T105243B).
Inspect, straighten and aside the pointer
4. Pry the dial from the case using a small flat head screwdriver.
5. Remove the movement from the case using T108897 and aside.
6. Loosen screw from the case using T105258 and remove the
capsule.
Reassemble In reverse Order
* - See section 8.5 for exploded view drawing for the Economy
Aneroid.