40
Program Mode
Welch Allyn TM286 Auto Tymp
Program Menu Page 2 Option Descriptions
DATA XFER CONFIG
These settings are used to allow the data transfer from the TM286 to a computer. The
settings must be the same on the TM286 and the computer. The factory defaults are
defined by an *.
POWER UP SETTINGS
The feature determines the mode displayed upon start up. An asterisk will denote which
option is selected to appear on the LCD when the system is first powered up. The factory
default setting is TYMP REFLEX.
PRN HEADER WA/PRN HEADER OFF/PRN HEADER CUSTOM
There are three options for the print header on the printout.
PRINT HEADER WA
This is the factory default setting for this feature. Each time the
Print Screen
or
All
tests in memory buttons are pressed, the printout will begin with the label
TM286.
PRINT HEADER OFF
If this option is selected, no header will be printed before any test results, which will save
space and printout time.
PRINT HEADER CUSTOM
Select this option to design a custom header, which might be the name of an individual
facility, department or company. To type in the custom header, position the square cursor
in front of
PRN HEADER CUSTOM
. Press
to select it as the new default setting.
The word
SAVED
appears in the lower right corner.
* 115.2 KBAUD
* NO 8-BIT
57.6 KBAUD
ODD 7-BIT
38.4 KBAUD
EVN 7-BIT
17.2
SPC 7-BIT
9600 BAUD
* XON/XOFF DISABLED
4800 BAUD
XON/XOFF ENABLED
TYMP
* TYMP REFLEX
AUDIO
M +
Summary of Contents for TM286
Page 18: ...14 Installation Welch Allyn TM286 Auto Tymp...
Page 36: ...32 Operation Welch Allyn TM286 Auto Tymp...
Page 60: ...56 Test Results Welch Allyn TM286 Auto Tymp...
Page 64: ...60 Computer Interface Welch Allyn TM286 Auto Tymp...
Page 72: ...68 Standards Specifications and Accessories Welch Allyn TM286 Auto Tymp...