19
Results
Reliability Number indicates the number of good readings obtained and their
consistency, based on a 1 to 9 scale. The higher the number, the better.
•
>
6 is typically acceptable
+
•
5 is marginal / repeat the test if possible
•
<
4 is poor / repetition of the test is necessary
+
If the patient did not appear to fixate on the unit during data acquisition, repeat
the test, even if reliability numbers are
>
6.
•
S
is the
sphere
in diopters.
•
C
is the
cylinder
power in diopters.
A +9.99 or -9.99 indicates a reading outside the unit’s measurement range.
•
The SW X.XXe after the date on the printout is the software revision level in
the SureSight unit.
Sphere
Cylinder
Reliablity
Number
Axis
Child/Adult
Mode