Configuring the Screen Results screen
WARNING
Patient Risk - selecting "hide recommendation" causes the
device to not show Pass/Refer messages during screening.
1. To select or change viewing options, touch
Tools
and then
Results
.
2. To select an option, touch the radio button next to it.
3. To return to the Tools screen, touch the
X
in the upper-left corner.
To screen in monocular mode
In monocular mode, you can choose which eye to screen. Use this mode when binocular
screening does not capture the subject’s pupils.
After a binocular screening completes:
1. At the end of the binocular screening process, touch
Mono
.
2. To select the subject's eye to screen, touch the corresponding side of the screen.
(OD = subject's right eye; OS = subject's left eye).
3. To repeat the process with the subject's other eye, touch
Mono
from the Subject
Results menu.
If a measurement is canceled or times out:
1. If a measurement times out after an unsuccessful binocular screening, touch
Mono
.
2. To select the subject's eye to screen, touch the corresponding side of the screen.
(OD = subject's right eye; OS = subject's left eye).
3. To repeat the process with the subject's other eye, touch
Mono
from the
Measurement Options menu.
Overview: Conducting a mass screening
If you set up a subject list ahead of time, you can screen several individuals quickly
without having to stop to add additional information.
Note
For best results, when the device has accumulated 300 records, delete all
records from the vision screener.
CAUTION
Verify you have successfully exported the records to a USB
flash drive before you touch Clear All. If you touch Clear All on the Queue
screen, all the unscreened subject information is deleted. If you touch Clear
All on the History screen, all completed exams on the device are deleted.
WARNING
Avoid using the vision screener adjacent to or stacked with
other equipment, including other vision screeners, or medical electrical
systems because doing so could result in improper operation. If such use is
necessary, observe the vision screener and other equipment to verify that
they are operating normally.
The following is an overview of the steps involved in screening multiple subjects.
1. Export the .csv files from the vision screener to a USB 2.0 drive. For more
information, see Import/Export.
Directions For Use
Screening subjects 23
Summary of Contents for Spot VS100
Page 1: ...Welch Allyn Spot Vision Screener Model VS100 Directions For Use Software version 3 1 XX ...
Page 18: ...14 Controls display window and connections Welch Allyn Spot Vision Screener Model VS100 ...
Page 22: ...18 Setting up the vision screener Welch Allyn Spot Vision Screener Model VS100 ...
Page 58: ...54 Specifications Welch Allyn Spot Vision Screener Model VS100 ...
Page 60: ...56 Warranty Welch Allyn Spot Vision Screener Model VS100 ...
Page 62: ...58 Service policy Welch Allyn Spot Vision Screener Model VS100 ...
Page 64: ...Material No 728900 ...