Software update messages
Message
Possible cause
Suggested action
Software Update:
Manifest transfer timed
out. Verify connection
and retry.
The manifest file transfer timed out
or connection was lost during
download
Verify the connection and retry.
Software Update:
Package file transfer
timed out. Verify
connection and retry
The package file transfer timed out
or connection was lost during
download
Verify the connection and retry.
Software Update: Invalid
token file.
There was an invalid token file
Verify and update the token file.
Software Update: Unable
to find manifest file on
server.
Unable to find manifest file on
server
Verify the manifest file is on the server.
Software Update: Unable
to verify manifest file
signature.
Unable to verify manifest file
signature
Regenerate the software package and retry.
Software Update:
Package file corrupted.
Regenerate package and
retry.
The package file is corrupt, does not
have expected SHA256 hash
Regenerate the software package and retry.
Software Update: Unable
to find package file.
The package file cannot be found
Verify the package file is on the server.
Software Update:
Installation failed. Reboot
and retry.
At least one of the sub systems
failed to install
Restart the device.
Software Update:
Upgrade unsuccessful.
Insufficient disk space.
The partition is running out of
space
Free up adequate space needed to perform
the upgrade.
Software Update: Update
unsuccessful.
Incompatible firmware.
The current firmware version is too
low to install update
Try to update to an earlier software package.
Software Update: SWUP
internal error
SWUP NIBP is not functional
SWUP internal error.
Software Update:
Manager internal error
The Software Update manager is
not functional
Software Update Manager internal error.
Identifying the device and subsystems
This explanation of the device labeling is provided to assist you in identifying these particular parts
by serial number. The part numbers appearing on the labels are internal part numbers and will not
be found in the list of field replaceable units (FRUs) included in this manual.
120 Appendix
Summary of Contents for Spot Vital Signs 4400
Page 1: ...Welch Allyn Spot Vital Signs 4400 Service manual Software version 1 X...
Page 8: ...4 Symbols and definitions...
Page 34: ...30 Power up sequence...
Page 68: ...64 Field replaceable units...
Page 127: ...Interconnect diagram Service manual Appendix 123...
Page 128: ...124 Appendix...
Page 132: ......