SP-2 User Guide
Art. No. 71022-0000 Welch Allyn Schiller ©2000
E
n
g
l
i
s
h
Calibration
Calibration Procedure
To calibrate the unit proceed as follows:
1.
Connect the calibration pump to the sensor. Ensure that there
are no air leaks.
2.
Press the up or down key (CALIB) to commence calibration.
3.
Enter the ambient temperature and press enter to confirm.
Temperature
:
21
o
C
Measured Vol.
:
0.00 l
Syringe Vol.
:
6.00 l
Deviation
:
-- %
hold sensor still
*** press START ***
MENU
4.
Wait 1 second.
5.
Press the START/STOP key .