Directions for Use
Mobile Stand (Optional)
31
Mounting the ProXenon Light Source
1.
Lock the caster wheels to prevent stand movement.
2.
Place the ProXenon 350 Surgical Illuminator (REF 902 series) on the platform so that
the rubber feet rest in the platform wells. Position the light source so that the front of
the light source is above the platform handle and the rear of the light source is above
the cord wrap.
3.
Secure the light source with two #10-32 captured thumbscrews. Wrap the power
cord around the cord wrap located at the rear of the platform.
Maintenance of the Mobile Stand
Cleaning
Treat the mobile stand as a non-patient contact environmental surface. Wipe with a
dampened cloth/wipe using a Quaternary-Amine/Low Alcohol Concentration Based
combination low or intermediate level cleaner-disinfectant that is labeled compatible with
metal, plastic, and painted surfaces.
Inspection
Periodically inspect the mobile stand for:
•
proper operation of casters and locking casters
•
secure connection between the pole to the base, platform, and hanger
•
secure connection between the lightsource and the platform
Repair or replace the mobile stand if any defects are found prior to use.
Specifications
Storage
-20° C to 49° C @ 95% relative humidity, max.
Platform
Height
915 mm
Base
Diameter
584 mm