Directions for Use
Care and Maintenance
19
4.
Grasp the handle and pull the entire lamp module out of the ProXenon 350 Surgical
Illuminator.
To insert a new lamp module:
1.
Orient the new lamp module so that the window of the Welch Allyn lamp REF 90209
is facing the front of the ProXenon 350 Surgical Illuminator.
2.
Align the sides of the lamp module and place it in the receiver channel. Firmly push in
the lamp module to ensure proper connection.
3.
Close and latch the door. If the door does not close completely, the lamp module is
not fully seated.
To Reset the Lamp Hours Meter:
1.
Verify that the lamp is off.
2.
Press and hold the
Full Dim +
Decrease
buttons.
3.
While holding the
Full Dim + Decrease
buttons
,
press and hold the
Lamp ON/OFF
buttons for three seconds. The
Lamp Hour
window flashes three times and resets to
zero.
Caution
Do not touch the face of the lamp (glass window).
Note
To cancel reset, release any button during the three-second period.
100-240V- 50HZ 6.0A INPUT
REF
06K1883D003
9020 0
Manufacutured For
XENON LIGHT SOURCE