8. To store data for later review or printing:
1. Makesure unit is ON.
2. Match pattern on button with observed ear.
3. Depress appropriate MEMORY
button.
4. To insure data entry into memory observe memory
symbol on screen.
CAUTION: Holding the MEMORY button down for more than
3 seconds will cause memory to be erased.
9.
To
start
new test, depress START/RUN
button.
10. Tips may be removed by hand or by sliding the
toward the tio.
white tip ejector
1 To
results, use Printer/Charger
(pg. 16).
12. To
display
right
or left
memory contents, press
appropriate memory button.
13. To
erase
right
or left
memory contents, depress
appropriate memory button continuously for 3 seconds. BLANK
MEMORY (-- symbol will appear.
14. The MicroTymp will turn OFF automatically 40 seconds after the
last test or last actuation of any button. Contents of memory are
not lost when the unit turns OFF.