• Mount unit by fitting key hole slots on back of unit over screw heads.
• Push unit against mounting plate while pushing down to lock screw heads into
small end of key hole slots.
S
ET
U
P
• Carefully remove plastic plug from light pipe holder on top of unit.
• Insert fiber optic light pipe into light pipe holder until it is secure.
• Plug power cord into base of unit and into AC grounded outlet.
• Position the power switch located on the top of the unit to ON.
N
OTE
:
Please refer to use of light pipe section
prior to using unit.
To ensure long unit life, high light out-
put and continued position rigidity, we
recommend that you do not acutely
bend the Light Pipe. For example: Middle
Bend—not more than 180°. End Bends—
not more than 90° each.
W
ARNING
:
Not for use in the presence of flammable anesthetic.
Do not use Model 48200 (Standard model) Light Pipe for neonate transillumination.
Even though the distal end of the light pipe does not feel warm to the touch, it can
burn neonates. Erythema may result from contact of more than 15 seconds. Use
Model 48210 Neonate Transilluminator for safe transillumination.
3
Plug
Light
Pipe
Max. 90°
Max.
180°