50
Chapter 5 Troubleshooting
Welch Allyn CP 200 Spirometry Option
Problem-Solving Suggestions
If you try these suggestions and still have problems, contact Welch Allyn. For phone
numbers, see
.
Add index entries for each error message, verbatim.
Condition
Causes
Actions
When printing, text prints correctly but
FVC curve does not.
Print settings
Make sure that the desired curve is selected. See
Unable to calibrate.
•
Poor connection between flow
transducer and sensor.
•
Damage to flow transducer.
•
Leak during calibration.
•
Uneven calibration strokes.
•
Check the connection between flow transducer
and sensor.
•
Replace the flow transducer if it is damaged.
•
Ensure that the connection between the
calibration syringe and flow transducer is tight
with no leaks.
•
Use even strokes in calibration.
No sensor detected.
Poor connection between the sensor and
the electrocardiograph.
Disconnect and reconnect the sensor.
Does not print.
•
Out of paper.
•
Paper jam.
•
Load paper. See the electrocardiograph manual.
•
If paper is jammed, clear it, then reload.
Values are too high (intermittent).
•
Patient’s fingers obstructed the screen
on the back of the flow transducer,
causing high back pressure and false
reading.
•
Patient’s lips were not tightly sealed
around the flow transducer.
•
Spirometer was calibrated with the
wrong size syringe.
•
Retest.
•
Recalibrate with a 3-liter syringe. See
Values are too high (consistently).
Pressure connection is partially obstructed. Remove any foreign substance from the flow
transducer or pressure tubing.
Predictive values are blank.
•
The selected norm does not support
certain values, and composite norm
values are disabled.
•
Re-enter age/birthdate, height, gender, race.
See
•
Enable composite norm values. See
The flow sensor has been dropped.
Accident.
Recalibrate. See
Report does not print parameters or
graphs.
Improper print settings.
Check print settings. See
Patient test values differ from values
expected by physician.
Various.
•
If the transducer is contaminated with sputum or
secretions, replace it.
•
Verify that proper barometric pressure has been
entered. See
•
Verify the patient data.
•
Eliminate any leaks in the pressure tubing.
•
Retest using a nose clip.
•
Replace the sensor if damaged.
•
Recalibrate.
•
Replace the transducer and retest.
Summary of Contents for CP 200
Page 1: ...CP 200 Spirometry Option Directions for Use ...
Page 34: ...30 Chapter 3 Calibrating the Spirometer Welch Allyn CP 200 Spirometry Option ...
Page 52: ...48 Chapter 4 Performing Spirometry Tests Welch Allyn CP 200 Spirometry Option ...
Page 53: ...49 5 Troubleshooting Problem Solving Suggestions 50 Limited Warranty 51 Service Policy 51 ...
Page 56: ...52 Chapter 5 Troubleshooting Welch Allyn CP 200 Spirometry Option ...
Page 80: ...76 Appendix D Predictive Norms etc Welch Allyn CP 200 Spirometry Option ...
Page 91: ......