Symptom
Possible cause
Suggested action
when high-speed printing has not
taken place.
The read-back file is corrupted.
During the Service USB Host test, the content
read back from file does not match the content
previously written.
Ensure that the thumb drive and
port are working properly. If
using another thumb drive results
in the same error, contact
Technical Support.
Internal printer is out of paper.
Reload.
The cue marker on the paper sensor is causing
the printer to incorrectly identify that paper tray
is empty.
Check Que Sensor board and
harness.
Post test error messages
Post test error messages:
Symptom
Possible cause
Suggested action
System Fault
code
01
Battery capacity is at critical capacity and auxiliary power is
not present
Ensure that AC power is plugged in.
Check for faulty battery and/or 60W power
supply.
Check all harness to power supply.
System Fault
code
02
Core image verification fails
Reload software through the Welch Allyn
Service Tool.
System Fault
code
03
Core image is missing
Reload software through the Welch Allyn
Service Tool.
System Fault
code
04
Software update failure
Reload software through the Welch Allyn
Service Tool.
System Fault
code
05
RAM Test failure
Check for faulty SMCE board.
66 Troubleshooting
CP 150 12-lead resting electrocardiograph
Summary of Contents for CP 150
Page 1: ...CP 150 12 lead resting electrocardiograph Service manual Software version 2 10 XX...
Page 8: ...4 Symbols CP 150 12 lead resting electrocardiograph...
Page 22: ...18 Overview CP 150 12 lead resting electrocardiograph...
Page 26: ...22 Controls indicators and connectors CP 150 12 lead resting electrocardiograph...
Page 27: ...Home screen Home screen The Home screen includes the following areas 23...
Page 30: ...26 Home screen CP 150 12 lead resting electrocardiograph...
Page 42: ...38 Settings CP 150 12 lead resting electrocardiograph...
Page 74: ...70 Guidance and manufacturer s declarations CP 150 12 lead resting electrocardiograph...
Page 81: ...Cable identification for reassembly Service manual Disassembly repair and reassembly 77...
Page 110: ...106 Disassembly repair and reassembly CP 150 12 lead resting electrocardiograph...
Page 128: ...124 Partners in Care service and support agreements CP 150 12 lead resting electrocardiograph...
Page 132: ...Interconnect diagram 128 Appendices CP 150 12 lead resting electrocardiograph...
Page 133: ......