Disassembly and Repair Procedures
Welch Allyn CP 100 and CP 200 Electrocardiograph
Replacing the Battery or Fuse
Procedure
1.
Turn the electrocardiograph up-side-down.
2.
Unscrew and remove the battery door. See
Figure 46. Removing the Battery Door
3.
Lift out the battery. See
.
Figure 47. Lifting Out the Battery
4.
Inspect the fuse. See
•
If the “Z” wire is intact, go to
.
•
If the “Z” wire is broken or dark,
replace the fuse. See
Figure 48. Battery Fuse Plugged In
Battery door
Battery
Fuse