When the 9600 plus and adapter temperature is stable, put the CareTemp thermometer in
calibration mode. Place the CareTemp in the adapter and press the measure button.
When the green LED flashes remove the CareTemp thermometer from adapter and read
thermometers temperature.
If the temperature is /- 0.2
⁰
C (+/- 0.4
⁰
F), of the Set Point the thermometer is within
calibration.
CareTemp 9600 Plus Adapter
Test Procedure