CareTemp Touch Free Thermometer
Service training
Page 1: ...CareTemp Touch Free Thermometer Service training...
Page 2: ...Topics We ll Cover Product and Accessories Reference documents Error messages Device Calibration Check Repair and technical...
Page 3: ...meter all regions 106091 410552 CareTemp No Touch Thermometer China 106090 410551 CareTemp DFU CD multi language Replacement Part 105802 410670 CareTemp Battery Door Replacement Part 106224 722755 QRC...
Page 4: ...eter 80018998 CareTemp Touch Free Thermometer DFU 106090 410551 CareTemp DFU CD multi language Covers How to use button functions taking measurement memory timer temp scale Trouble shooting and error...
Page 5: ...y less than 10 Suggest Replace battery Cause The measurement is lower than 93 2 F 34 C Suggest Retake measurement following procedure Cause The measurement is higher than 109 4 F 43 C Suggest Retake m...
Page 6: ...DEVICE MAINTENANCE CALIBRATION CHECK...
Page 7: ...9600 plus calibration tester at all three set point levels across the temperature range The adapter is attached externally to the 9600 plus allowing customers to install without voiding the warranty...
Page 8: ...Temp 9600 Plus Adapter you will need the following CareTemp 9600 Plus Adapter Parts 9600 plus calibration tester 9600 power supply 9600 Plus CareTemp Adapter Power cable Mounting bracket Two screws Ph...
Page 9: ...the 9600 plus calibration tester using the mounting bracket and screws Use the power cable to connect the adapters DC out connection to the 9600 plus power connector Connect the power supply to the a...
Page 10: ...er up the 9600 plus and adjust the set point The adapters temperature LED turn on and the Ready LED flashes until the temperature stabilizes within 05 C of the set point CareTemp 9600 Plus Adapter Set...
Page 11: ...CareTemp 9600 Plus Adapter Startup Wait for the temperature to stabilize as indicated by 9600 Plus LCD stops flashing Adapters Ready LED stops flashing 0 25 C of Set Point...
Page 12: ...sure 4 Memory 5 C F Celsius Fahrenheit 6 Pulse timer A With the thermometer off press MEM button 4 and timer button 6 at the same time and hold for 3 seconds The display will show the test screen then...
Page 13: ...us Adapter Test Procedure The thermometer will remain in calibration mode for 60 seconds before shutting off Wait at least 1 minute between temperature measurements Measurements performed too frequent...
Page 14: ...ace the CareTemp in the adapter and press the measure button When the green LED flashes remove the CareTemp thermometer from adapter and read thermometers temperature If the temperature is within 0 2...
Page 15: ...ibration tester set the 9600 Plus to the next Set Point and wait for the temperature to stabilize before repeating steps 1 to 3 Wait at least 1 minute between temperature measurements Measurements per...
Page 16: ...nezuela Brazil is through Authorized Service Provider Welch Allyn Mexico supporting Mexico Welch Allyn Tilburg Netherlands Supporting Europe and India Welch Allyn UAE Supporting the Middle East Welch...
Page 17: ...eTemp 9600 Plus Adapter setup CareTemp 9600 Plus Adapter test procedure WA website Product page Link to technical support The 9600 Plus calibration tester manual will detail the procedure for verifyin...
Page 18: ...ive device to nearest Welch Allyn center is not included and is the customer s responsibility US For OOBF or customer satisfaction you may offer a shipping label International Shipping cost from locat...
Page 19: ...ted to determine if the device has failed using the procedure detailed in the 9600 Plus calibration tester manual The original warranty will remain in effect from the date of original purchase to the...
Page 20: ...QUESTIONS AND DISCUSSION...