4 Preparation and operation
Pausing ventilation in CPR CCSV mode
During the analysis of the defibrillator, you can pause ventilation
in order to avoid ventilation artifacts in the analysis.
Requirement
•
The device is switched on
•
The CPR CCSV mode is set.
1. Pause chest compressions.
Result
The ventilation is paused for as long as the chest compressions
are continued (up to the set back-up ventilation time at most).
4.7.12 Changing the ventilation mode in CPR mode
This chapter describes how you can switch between CPR Manual,
CPR IPPV and CPR CCSV in CPR mode.
Requirement
•
The device is switched on.
•
A CPR mode is set.
If the trigger is still not functioning properly in
CPR CCSV
mode
in spite of reducing the trigger and increasing the PEEP, you can
switch from
CPR CCSV
mode to
CPR IPPV
mode
(see "4.7.12 Changing the ventilation mode in CPR mode",
page 84)
.