5. The oxygen must be felt to come out at the
breathing station.
6. Switch off inhalation immediately afterwards to
save oxygen.
12.5 Checking aspiration
Required for Suction
MODULE and Combi
MODULE
1. Open the oxygen supply valve
slowly
.
2. Switch on aspiration.
3.
Suction MODULE:
turn the dial to the right-hand
stop. Take off the vacuum tube and hold
vacuum connection
closed.
Combi MODULE:
the vacuum is permanently set to
≥
– 0.5 bar (at 4.5 bar O
2
inlet pressure) and can-
not be read off directly on the pressure gauge. You
should therefore plug WEINMANN Emergency test
pressure gauge WM 15294 onto connection for
the vacuum
.
4. Read off the value which becomes constant on
the pressure gauge (Suction MODULE) or on the
test pressure gauge (Combi MODULE). This value
must be at least – 0.5 bar at a feed pressure of
4.5 bar.
5.
Suction MODULE:
to check that the OMNIVAC
secretion aspiration system is leaktight, connect
the vacuum tube again and hold the aspiration
opening of the end-piece closed instead.
Combi MODULE:
to check that the OMNIVAC
secretion aspiration system is leaktight, connect
WEINMANN Emergency test pressure gauge
WM 15294 to the end-piece of the fingertip
regulator. Hold the fingertip regulator closed as
you do so.
6. After no more than 30 seconds, at least –
0.5 bar must be displayed.