Hygienic preparation
29
EN
5. Check whether the pressure measurement tube is still correctly fitted in the up-
per part. If necessary reinsert it.
6. Fit the upper and lower parts together such that
the marks are aligned. Turn the upper and lower
parts against each other clockwise as far as they
will go.
7. Remove the closure. The humidifier is now ready
for use again.
For hygiene reasons, we recommend that the plastic parts are replaced after a
maximum of two years use.
CLICK 2 drying
If water penetrates into the pressure line, then dry CLICK 2 as follows:
1. Push the bellows included in the delivery, as illus-
trated, onto the end of the pressure line at the
outlet connectors.
Caution!
The bellows must
only
be used for drying the humidifier. During drying
CLICK 2 must not be connected to the therapy unit. Otherwise the unit
could be damaged.
2. Press the bellows together so that any moisture is blown out.
Caution!
Keep the bellows pressed in so that moisture is not sucked back in again.
3. Remove the bellows.
4. Repeat steps
two or three times.
Marks