46
Repairguide for VENTImotion/VENTIlogic
7.13 Replace bottom part of housing
Tools and equipment required
• ESD protected work-station
• 3 mm Allen key
• Phillips screwdriver, size 1
• Torque wrench 20 – 120 Ncm
1. Open the device (see section 7.2 on
2. Undo screws
the ferrite core.
3. Pull connecting cable
for the humidifier out
upwards.
4. Undo screws
and remove the alarm (piezo
buzzer).
5. Remove the foam for the expansion space.
6. Remove the plug locators for the O
2
valve.
You have now dismantled all parts. Now start as-
sembling into the new bottom part of the housing
7. Insert the plug locators for the O
2
valve in the
new bottom part of the housing.
8. Put the foam for the expansion space back in.
9. Refit the alarm (piezo buzzer) and fix it in posi-
tion with the two screws
Tighten the screws
with a torque of 47 Ncm.
10. Put connecting cable
into the plug locator
with the radius of the plug side facing
down-
wards
. Both webs of the plug must be located
within
the plug locator for this.
11. Fix connecting cable
washer
.
12. Close the device (see section 7.3 on
13. Test the device (see "4. Test the device" on
Incorrect:
Correct: