Repairguide for VENTImotion/VENTIlogic
45
7.12 Replace fascia film
Tools and equipment required
• ESD protected work-station
• 3 mm Allen key
• Phillips screwdriver, size 1
• Knife with a smooth, flat blade
1. Open the device (see section 7.2 on
2. Remove the mains supply unit (see section 7.4
3. Remove the power board (see section 7.5 on
4. Remove box and filter holder (see section 7.6
5. Remove the control board and the display (see
section 7.8 on page 39). Undo fascia film
using a knife-blade and carefully peel it off.
Use a little 70 % isopropanol to degrease this
part of the housing.
6. Peel the protective film off the inside of the new
fascia film and affix fascia film
to this area
of the housing. Ensure that the ribbon cable is
pushed cleanly through the opening in the hous-
ing without any kinks.
7. Carefully pull the outer protective film off the
new fascia film using a fingernail.
8. Refit the display and the control board (see
9. Refit the box and filter holder (see section 7.6
10. Refit the power board (see section 7.5 on
11. Refit the mains supply unit (see section 7.4 on
12. Close the device (see section 7.3 on